| Catalog Number |
81251170S |
| Medical Device Problem Codes |
Positioning Failure (1158); Device Damaged by Another Device (2915)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/27/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Once the investigation has been completed any additional information will be reported in a supplemental report.
|
| |
|
Event or Problem Description
|
|
As reported: "customer was not able to hit the nail correctly with the femoral neck screw drill.The nail was hit with the drill, this also happened with the second nail.The nail holding screws show signs of wear.".
|
| |
|
Additional Manufacturer Narrative
|
|
Please note the correction to h6 health impact codes.The reported event could be confirmed during the investigations.The device inspection revealed the following: the identification of the returned nail could be confirmed based on the catalog and lot number.Signs of misaligned drilling can be observed on the inner part of the lag screw hole, confirming the reported event.The returned nail, lag screw reamer and both nail holding screws were tested with fully functional targeting device, targeting sleeve and tissue protection sleeve samples.With both nail holding screws, the assembling could be performed as required and the reamer could be perfectly aligned with the lag screw hole.The observations from the functional inspection confirm that the devices are fully functional.Based on investigation, the root cause was attributed to an user related issue.The functional test enabled to rule out any device related issue.A review of the device history for the reported lot did not indicate any abnormalities.No corrective actions are required at this time.A review of the labeling did not indicate any abnormalities.No indications of material, manufacturing or design related problems were found during the investigation.If more information is provided, the case will be reassessed.
|
| |
|
Event or Problem Description
|
|
As reported: "customer was not able to hit the nail correctly with the femoral neck screw drill.The nail was hit with the drill, this also happened with the second nail.The nail holding screws show signs of wear.".
|
| |
|