| Model Number |
101-9810 |
| Medical Device Problem Code |
Material Integrity Problem (2978)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/10/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that during a superion implant procedure, the washer on the spacer stripped and broke.Thick bone was noted and the physician used more force to try deploying the implant due to the osteophyte.The procedure was completed with a different spacer.The broken spacer was not returned to bsc for analysis as it was discarded by the medical facility.
|
| |
|
Additional Manufacturer Narrative
|
|
A retroactive review of events within the quality system identified this record as requiring remediation, including reporting and submission of an initial emdr per applicable criteria.Boston scientific has opened a corrective and preventive action (capa 8647) investigation to determine the root cause for why an initial emdr was not previously submitted to the fda in accordance with boston scientific established mdr reporting process and fda adverse event reporting requirements.Information gained through these investigation efforts will be utilized to determine root cause and implement appropriate preventive action(s), as necessary.
|
| |
|
Search Alerts/Recalls
|