|
Updated fields: b4, d9, g3, g6, h2,h3, h6(investigation findings, type of investigation, investigation conclusions,component code), h11.Corrected field : b5 ,d4 , e1 ( event site address , event site telephone , initial reporter ).A getinge field service engineer (fse) evaluated the unit and found the ecg connector to be broken.The fse replaced the faulty part: cbl assy ecg input to frnt end (0012-00-0976).All functional and safety checks have been performed to meet factory specifications.The device was return to customer for clinical use.
|