| Model Number |
M00558350 |
| Medical Device Problem Codes |
Burst Container or Vessel (1074); Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/29/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Block h6: imdrf device code a0402 captures the reportable event of a balloon burst.
|
| |
|
Event or Problem Description
|
|
It was reported to boston scientific corporation that a cre fixed wire dilatation balloon was attempted to be used in the distal esophagus during a dilation procedure to treat esophageal stricture performed on (b)(6) 2025.During the procedure, the balloon burst at 11mm during balloon inflation.It was reported that no piece of the balloon detached inside the patient.The procedure was completed with another cre fixed wire dilatation balloon.There were no patient complications reported as a result of this event.
|
| |
|
Additional Manufacturer Narrative
|
|
Block h6: imdrf device code a0402 captures the reportable event of a balloon burst.Block h11: investigation results: the returned cre fixed wire dilatation balloon was analyzed, and a visual inspection found that the balloon was longitudinally torn.Microscopic inspection confirmed the balloon longitudinal tear.No other problems with the device were noted.With all the available information, boston scientific concludes the reported event of balloon burst was not confirmed.The results of the analysis performed on the returned device found that the balloon had a longitudinal tear.It is possible that the longitudinally torn found in the balloon occurred due to procedural factors such as excess of pressure or interaction with other devices.Also, it is possible that interaction with a sharp surface during/previous the procedure, could have caused the problem found on the balloon.Therefore, the most probable root cause is adverse event related to procedure.
|
| |
|
Event or Problem Description
|
|
It was reported to boston scientific corporation that a cre fixed wire dilatation balloon was attempted to be used in the distal esophagus during a dilation procedure to treat esophageal stricture performed on (b)(6) 2025.During the procedure, the balloon burst at 11mm during balloon inflation.It was reported that no piece of the balloon detached inside the patient.The procedure was completed with another cre fixed wire dilatation balloon.There were no patient complications reported as a result of this event.
|
| |
|