DEPUY IRELAND - 9616671 ATTUNE SHIM SZ7 7MM; TIBIAL INSERT TRIALS Back to Search Results
Catalog Number 254500744
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/07/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Product complaint #: (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Event or Problem Description
It was reported that during the removal of the trial poly, the 7rp artic surf broke on the bottom of the instrument.A small piece broke off and was removed from the sterile field.The 7mm size 7-8 shim also broke into 2 pieces down the middle.Both pieces were removed from the sterile field.Surgery time was not extended.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Investigation summary: according to the information received that "it was reported that during the removal of the trial poly, the 7rp artic surf broke on the bottom of the instrument.A small piece broke off and was removed from the sterile field.The 7mm size 7-8 shim also broke into 2 pieces down the middle.Both pieces were removed from the sterile field.Surgery time was not extended.¿ the product was returned to j&j medtech orthopaedics for evaluation.Visual analysis of the returned device revealed that the attune shim sz7 7mm was broken into two pieces.Fragment was returned for evaluation.No other anomalies were noted.The failure mode is consistent with using a device in a prying motion to disassemble the mating instruments after trailing resulting in material overload at the smallest cross-sectional area in middle of the shim.No material or manufacturing defects were observed that would have contributed to the fracture.A functional test was not performed since it was not applicable to the complaint condition.A dimensional inspection was not performed since it was not applicable to the complaint condition.The overall complaint was confirmed as the observed condition of the attune shim sz7 7mm would have contributed to the complained issue.Based on the investigation findings, the potential cause is traced to unintended use error, and it has been determined that no corrective and/or preventative action is proposed.There is no indication that a design or manufacturing issue has caused the reported complaint condition.As part of j&j medtech orthopaedics quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device history lot: the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation (mre) was not performed.Corrected: h3.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ATTUNE SHIM SZ7 7MM
Common Device Name
TIBIAL INSERT TRIALS
Manufacturer (Section D)
DEPUY IRELAND - 9616671
loughbeg ringaskiddy co.
cork
EI 
Manufacturer Contact
elba bello
700 orthopaedic dr.
warsaw, IN 46581
9083863534
MDR Report Key23391411
Report Number1818910-2025-18317
Device Sequence Number10052666
Product Code HWT
UDI-Device Identifier10603295134169
UDI-Public(01)10603295134169(10)MVMBLC940
Combination Product (Y/N)N
Initial Reporter StateGA
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number254500744
Device Lot NumberMVMBLC940
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/21/2025
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 10/07/2025
Supplement Date Received by Manufacturer10/27/2025
Initial Report FDA Received Date10/27/2025
Supplement Report FDA Received Date10/28/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured03/23/2021
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexMale
-
-