| Catalog Number |
72290038 |
| Medical Device Problem Codes |
Overheating of Device (1437); Electrical Shorting (2926)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/29/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that during an acromioplasty, two (2) werewolf flow 90 wands overheated and reached temperatures of 56¿60 degrees celsius, even though the product was in the air, not obstructed, and placed in the pouch without suction.The same error message appeared with two different handpieces, indicating a ¿short circuit in the wand.¿ the procedure was completed after a surgical delay of approximately 10 minutes using a similar sn backup device.No further complications were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
H11: internal complaint reference: (b)(4).H3, h6: this complaint was opened by smith+nephew to document a product problem associated with a smith+nephew device.The reported problem relates to known inherent device and/or procedural risks that are appropriately documented in our risk files.Smith+nephew will continue to monitor trends in accordance with our post-market surveillance process and take necessary action as required if anticipated severity and/or occurrence rates are exceeded.Smith+nephew has no reason to suspect that the product failed to meet any specifications at the time of manufacture.Based on our review of all currently available information, we are unable to identify a definitive root cause.However, as the use of our product cannot be excluded as a potential cause or contributory factor to the reported issue, we are conservatively submitting this report in accordance with applicable regulations.If additional investigative findings or information becomes available that alters the conclusions of this report, a follow-up report will be submitted as required.
|
| |
|
Additional Manufacturer Narrative
|
|
Internal complaint reference case- (b)(4).H11, h3, h6: the information received by the manufacturer has been re-evaluated for mdr reporting and it was determined that this case does not meet the threshold for reporting and is a non-reportable event.The reported temperature of 56¿60 degrees celsius is under the normal temperature that the wand can function without overheating.If further details are provided confirming the occurrence of a reportable event, our files will be updated accordingly, and a further report submitted outlining both the event details and our investigations performed.
|
| |
|
Search Alerts/Recalls
|