| Model Number |
DTMB1D1 |
| Medical Device Problem Codes |
Failure to Charge (1085); Delayed Charge Time (2586)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/17/2020
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that the right atrial (ra) lead exhibited undersensing triggering device classified false termination of atrial tachycardia/atrial fibrillation (at/af) event(s) and inappropriate atrial pacing at times.It was also noted that all atrial tachycardia/atrial fibrillation (at/af) therapies had been disabled due to an atrial lead position check failure.In addition, a charge circuit timeout had occurred as the cardiac resynchronization therapy defibrillator (crt-d) was at end of service (eos).The device was programmed off.The crt-d and ra lead remain implanted.No patient complications have been reported as a result of this event.
|
| |
|
Additional Manufacturer Narrative
|
|
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Additional Manufacturer Narrative
|
|
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Event or Problem Description
|
|
It was further reported that the patient presented to the emergency department and is now home on hospice.The patient has declined any additional interventions.
|
| |
|
Event or Problem Description
|
|
It was reported that the right atrial (ra) lead exhibited undersensing triggering device classified false termination of atrial tach ycardia/atrial fibrillation (at/af) event(s) and inappropriate atrial pacing at times.It was also noted that all atrial tachycardia/atrial fibrillation (at/af) therapies had been disabled due to an atrial lead position check failure.In addition; a charge circuit timeout had occurred as the cardiac resynchronization therapy defibrillator (crt-d) was at end of service (eos).It was also noted that the device triggered an alert for long charge time post recommended replacement time (rrt).The device was programmed off.The crt-d and ra lead remain implanted.No patient complications have been reported as a result of this event.It was further reported that the patient presented to the emergency department and is now home on hospice.The patient has declined any additional interventions.
|
| |
|
Additional Manufacturer Narrative
|
|
Correction: b5, annex a.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Search Alerts/Recalls
|