ZIMMER ORTHOPAEDIC MFG. LTD. PERSONA FEMUR CEMENTED (CR) NARROW LEFT SIZE 9; PROSTHESIS, KNEE Back to Search Results
Catalog Number 42502006601
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Joint Laxity (4526)
Date of Event 09/29/2025
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).D2, d4, g4: diligence is in process to provide product identification information; however, no further information has been provided at this time.D10: ---------------- unk patella lot# unk 42512100513 articular surface medial congruent (mc) left 13 mm, lot # 65541183.42532006701 tibia cemented 5 degree stemmed left size d, lot # 66360624.42557000114 stem extension tapered cemented 14mm dia 30mm, l lot # 66060290.G2: australia.H6: proposed component code: mechanical (g04) - femur.The product will not be returned to zimmer biomet for investigation, as the device location is unknown.The investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.
 
Event or Problem Description
It was reported that there was a revision tkr due to patella instability/maltracting.Due diligence is in process for this complaint.To date no additional information has been received.
 
Event or Problem Description
No further event information available at the time of this report.
 
Additional Manufacturer Narrative
This report is being submitted to provide additional information.Updated: b4, b5, d2b, d10, g1, g3, g6, h1, h2, h3, h6, h10, and h11 no product was returned, or pictures provided; therefore, visual and dimensional evaluations could not be performed.Medical records were not provided.Review of the device history records identified no deviations or anomalies during manufacturing related to the reported event.A definitive root cause cannot be determined.This complaint is not confirmed.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PERSONA FEMUR CEMENTED (CR) NARROW LEFT SIZE 9
Common Device Name
PROSTHESIS, KNEE
Manufacturer (Section D)
ZIMMER ORTHOPAEDIC MFG. LTD.
building no 2 east park
shannon industrial estate
shannon, county clare
Manufacturer (Section G)
ZIMMER ORTHOPAEDIC MFG. LTD.
building no 2 east park
shannon industrial estate
shannon, county clare
Manufacturer Contact
jennifer rapsavage
56 e. bell dr.
warsaw, IN 46582
5745260384
MDR Report Key23393001
Report Number3007963827-2025-00421
Device Sequence Number10015684
Product Code MBH
UDI-Device Identifier00889024229952
UDI-Public(01)00889024229952(17)330925(10)66055831
Combination Product (Y/N)N
Initial Reporter CountryAS
PMA/510(K) Number
K172524
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2024
Device Explanted Year2025
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Distributor
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 12/29/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Lay User/Patient
Device Catalogue Number42502006601
Device Lot Number66055831
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/29/2025
Supplement Date Received by Manufacturer12/03/2025
Initial Report FDA Received Date10/27/2025
Supplement Report FDA Received Date12/29/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured10/03/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient SexUnknown
-
-