| Model Number |
FP7 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
Intraocular Pressure Decreased (4468)
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The ifu for the reported product model was reviewed and confirmed to include hypotony as a known complication.The device history record for the reported lot was reviewed and no issues were observed.Product was manufactured, tested, and released per validated procedures.The device was reported to be explanted and pending return to new world medical.Upon device return and evaluation, an addendum will be filed.
|
| |
|
Event or Problem Description
|
|
Ahmed valve was inserted but severe hypotony noted which could not be stabilised.Valve needed to be removed and replaced with a new ahmed valve.The second valve appears to be draining as expected immediately post operative but the patient will be followed up tomorrow.Due to extensive length of surgery patient needed an unplanned overnight stay.
|
| |
|
Additional Manufacturer Narrative
|
|
The explanted valve underwent visual inspection and functional testing.No issues were found with the device during visual inspection.A steady state pressure test was also conducted on the unit to mimic the physiological flow conditions of the human eye, where the resulting pressure in the system measured and yielded passing results.The returned valve met the acceptance criteria and performed as expected.The issue of low iop as reported by customer could not be replicated or confirmed.
|
| |
|