• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: NEW WORLD MEDICAL, INC. AHMED GLAUCOMA VALVE; GLAUCOMA DRAINAGE DEVICE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

NEW WORLD MEDICAL, INC. AHMED GLAUCOMA VALVE; GLAUCOMA DRAINAGE DEVICE Back to Search Results
Model Number FP7
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Intraocular Pressure Decreased (4468)
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The ifu for the reported product model was reviewed and confirmed to include hypotony as a known complication.The device history record for the reported lot was reviewed and no issues were observed.Product was manufactured, tested, and released per validated procedures.The device was reported to be explanted and pending return to new world medical.Upon device return and evaluation, an addendum will be filed.
 
Event or Problem Description
Ahmed valve was inserted but severe hypotony noted which could not be stabilised.Valve needed to be removed and replaced with a new ahmed valve.The second valve appears to be draining as expected immediately post operative but the patient will be followed up tomorrow.Due to extensive length of surgery patient needed an unplanned overnight stay.
 
Additional Manufacturer Narrative
The explanted valve underwent visual inspection and functional testing.No issues were found with the device during visual inspection.A steady state pressure test was also conducted on the unit to mimic the physiological flow conditions of the human eye, where the resulting pressure in the system measured and yielded passing results.The returned valve met the acceptance criteria and performed as expected.The issue of low iop as reported by customer could not be replicated or confirmed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AHMED GLAUCOMA VALVE
Common Device Name
GLAUCOMA DRAINAGE DEVICE
Manufacturer (Section D)
NEW WORLD MEDICAL, INC.
10763 edison court
rancho cucamonga CA 91730
Manufacturer (Section G)
NEW WORLD MEDICAL, INC.
10763 edison court
rancho cucamonga CA 91730
Manufacturer Contact
cristina avalos
10763 edison court
rancho cucamonga, CA 91730
9094664304
MDR Report Key23393150
Report Number1000125279-2025-00039
Device Sequence Number15913750
Product Code KYF
UDI-Device Identifier00892064002119
UDI-Public(01)00892064002119(11)250131(17)270131(10)A1125
Combination Product (Y/N)N
Initial Reporter CountryUK
PMA/510(K) Number
K230975
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Distributor
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 11/20/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model NumberFP7
Device Catalogue NumberFP7
Device Lot NumberA1125
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/30/2025
Supplement Date Received by ManufacturerNot provided
Initial Report FDA Received Date10/27/2025
Supplement Report FDA Received Date11/24/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured01/31/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age84 YR
Patient SexUnknown
-
-