Catalog Number 1125000030
Medical Device Problem Code Electrical /Electronic Property Problem (1198)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Event or Problem Description
This record is a consolidation of records summarized as a part of the fda voluntary malfunction summary reporting program.Events occurred between july 1-september 30, 2025.This report summarizes 2 malfunction events(s), where it was reported the devices experienced accessible ac current.No adverse consequence or clinically relevant delay in treatment was reported.
 
Additional Manufacturer Narrative
This record is a consolidation of records summarized as part of the fda voluntary malfunction summary reporting program.2 device(s) was functionally/visually inspected in the field.The device(s) was/were repaired and returned to use.Evaluation results findings (components/results): 2 devices: power cord / current leakage problem.There was no remedial action taken.This device is not labeled for single use.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PRIME ZOOM STRETCHER30INLITTER
Common Device Name
STRETCHER, WHEELED, POWERED
Manufacturer (Section D)
STRYKER MEDICAL-KALAMAZOO
3800 east centre avenue
portage MI 49002
Manufacturer (Section G)
STRYKER MEDICAL-KALAMAZOO
3800 east centre avenue
portage MI 49002
Manufacturer Contact
melissa simon
3800 east centre avenue
portage, MI 49002
2693292100
MDR Report Key23393435
Report Number0001831750-2025-99387
Device Sequence Number13273688
Product Code INK
UDI-Device Identifier07613327278293
UDI-Public7613327278293
Combination Product (Y/N)N
Initial Reporter StateMI
Initial Reporter CountryUS
Exemption NumberVMSR
Number of Events Summarized2
Summary Report (Y/N)Y
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 10/27/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Catalogue Number1125000030
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 10/01/2025
Initial Report FDA Received Date10/27/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1