BOSTON SCIENTIFIC CORPORATION ENDOTAK RELIANCE; IMPLANTABLE LEAD Back to Search Results
Model Number 0266
Medical Device Problem Codes Signal Artifact/Noise (1036); Over-Sensing (1438); Pacing Problem (1439)
Health Effect - Clinical Code Asystole (4442)
Date of Event 08/22/2025
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that during a follow up visit, this cardiac resynchronization therapy defibrillator (crt-d) system exhibited noise on the right ventricular (rv) lead.It was noted that the patient was recently implanted with an additional concomitant device.Technical services (ts) was called fora consult review of the presenting electrogram (egm).Upon review, the presenting egm showed there to be 3.5 seconds of asystole that appeared to have been caused by oversensed myopotential noise.The health care professional (hcp) stated that the noise was reproduced in the clinic.Ts discussed programming optimization options with the hcp.No adverse patient effects were reported.At this time this crt-d and rv lead remain in service.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENDOTAK RELIANCE
Common Device Name
IMPLANTABLE LEAD
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
4100 hamline avenue north
saint paul MN 55112
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
road 698, lot no. 12
dorado PR 00646 -260
*   00646-2602
Manufacturer Contact
timothy degroot
4100 hamline avenue north
saint paul, MN 55112
6515826168
MDR Report Key23393445
Report Number2124215-2025-76975
Device Sequence Number15920643
Product Code NVY
UDI-Device Identifier00802526431296
UDI-Public(01)00802526431296(17)210219
Combination Product (Y/N)Y
Initial Reporter StateGA
Initial Reporter CountryUS
PMA/510(K) Number
P910073
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2020
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 10/27/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date02/19/2021
Device Model Number0266
Device Catalogue Number0266
Device Lot Number420045
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/30/2025
Initial Report FDA Received Date10/27/2025
Date Device Manufactured02/19/2019
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age77 YR
Patient SexMale
-
-