• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY ORTHOPAEDICS INC US PFC*SIGMA C/R NPOR FEM LT SZ5; SIGMA KNEE PRIMARY : KNEE FEMORAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY ORTHOPAEDICS INC US PFC*SIGMA C/R NPOR FEM LT SZ5; SIGMA KNEE PRIMARY : KNEE FEMORAL Back to Search Results
Catalog Number 960005
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Pain (1994); Joint Laxity (4526)
Date of Event 10/08/2025
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Event or Problem Description
It was reported that the patient underwent revision for pain & instability.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information that has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint recon described in this report.H11 additional narrative: added: h6.Health effect - clinical code.If information that was not available for the initial report is obtained, a follow-up report will be filed as appropriate.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Investigation summary: the attached were revised yesterday, reason for revision was pain & instability.Surgeon does not require feedback.The product was not returned to depuy synthes; however, photos were provided for review.The x-rays provided were reviewed.However, with the provided information no signs of a device nonconformance were observed.As the device was not returned, an as-received condition could not be assessed, and a dimensional inspection and document/specification review were not completed.The overall complaint was not confirmed as the observed condition of the pfc*sigma c/r npor fem lt sz5 would have not contributed to the reported adverse event.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.There is no indication that a design or manufacturing issue has caused the reported complaint condition.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device history lot: the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation (mre) was not performed.H11.Additional narrative: added: d10 (concomitant).Corrected: h4.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Common Device Name
SIGMA KNEE PRIMARY : KNEE FEMORAL
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
Manufacturer (Section G)
CORK MFG & MATERIAL WAREHOUSE
loughbeg ringaskiddy
cork
EI  
Manufacturer Contact
elba bello
700 orthopaedic dr.
warsaw, IN 46581
9083863534
MDR Report Key23393574
Report Number1818910-2025-18350
Device Sequence Number21377182
Product Code JWH
UDI-Device Identifier10603295231974
UDI-Public(01)10603295231974(17)240531(10)9194170
Combination Product (Y/N)N
Initial Reporter CountryUK
PMA/510(K) Number
K943462
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2020
Device Explanted Year2025
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date05/31/2024
Device Catalogue Number960005
Device Lot Number9194170
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 10/08/2025
Supplement Date Received by Manufacturer12/23/2025
01/05/2026
Initial Report FDA Received Date10/27/2025
Supplement Report FDA Received Date12/26/2025
01/06/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured06/20/2019
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
PFC*SIGMA/OV/DOME PAT 3PEG,38; SIG MOD TIB TRAY CEM COCR 4; SIGMA PLI GVF INS 4 10MM
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-