| Catalog Number |
256066 |
| Medical Device Problem Code |
False Positive Result (1227)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/05/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported while using the bd veritor¿ plus analyzer, one (1) false positive flu b patient result was obtained, simultaneously with a positive sars-cov-2 result.Upon confirmatory testing using a standalone flu test cartridge, the patient sample was negative for flu b.There were no health impact or consequences reported.
|
| |
|
Additional Manufacturer Narrative
|
|
A device evaluation is anticipated but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.
|
| |
|
Additional Manufacturer Narrative
|
|
The following fields have been updated with additional information: d9.Device available for evaluation: yes.D9.Returned to manufacturer on: 11-nov-2025.H3.Device eval by manufacturer? yes.Investigation summary: the complaint alleges the bd veritor plus analyzer is having false positives for triplex assay (catalog number 256066, serial number (b)(6)).A replacement analyer was provided to the customer.Bd veritor plus analyzer was returned for investigation.Based on the test results in the csv file, it was showing positive result for flu b which was consistent with what was displayed on the screen.However, without access to the tiff images, we are unable to determine the root cause.Therefore, this complaint is confirmed.Device history record review for veritor plus analyzer, serial number (b)(6) was reviewed and nonconformances related to this failure mode were not revealed in manufacturing during final testing.The device was released conforming.Review of risk management files confirms there are no new or modified risks associated with this failure mode.Bd quality will continue to monitor for trends associated with the failure mode described in this complaint.
|
| |
|
Event or Problem Description
|
|
It was reported while using the bd veritor¿ plus analyzer, one (1) false positive flu b patient result was obtained, simultaneously with a positive sars-cov-2 result.Upon confirmatory testing using a standalone flu test cartridge, the patient sample was negative for flu b.There were no health impact or consequences reported.
|
| |
|
Additional Manufacturer Narrative
|
|
The requested 510k information for the involved veritor assay is as follows:
g4.Pma/510(k)#: eua(b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported while using the bd veritor¿ plus analyzer, one (1) false positive flu b patient result was obtained, simultaneously with a positive sars-cov-2 result.Upon confirmatory testing using a standalone flu test cartridge, the patient sample was negative for flu b.There were no health impact or consequences reported.
|
| |
|