| Catalog Number |
1115000030 |
| Medical Device Problem Code |
Mechanical Jam (2983)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
This record is a consolidation of records summarized as a part of the fda voluntary malfunction summary reporting program.Events occurred between july 1-september 30, 2025.This report summarizes 6 malfunction events(s), where it was reported the devices experienced steer mechanism is difficult to engage/disengage.No adverse consequence or clinically relevant delay in treatment was reported.
|
| |
|
Additional Manufacturer Narrative
|
|
This record is a consolidation of records summarized as part of the fda voluntary malfunction summary reporting program.6 device(s) was functionally/visually inspected in the field.The device(s) was/were repaired and returned to use.Evaluation results findings (components/results): 4 devices: rod/shaft / mechanical problem identified.1 device: chassis/frame / device migration.1 device: bolt / device migration.There was no remedial action taken.This device is not labeled for single use.
|
| |
|