| Catalog Number |
2141PX2000 |
| Medical Device Problem Code |
Unintended System Motion (1430)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
This record is a consolidation of records summarized as part of the fda voluntary malfunction summary reporting program.1 device(s) was functionally/visually inspected in the field.The device(s) was/were repaired and returned to use.1 device was not evaluated, as a probable cause for the issue was identified during communication between the customer and stryker and no further assistance was requested by the customer.Evaluation results findings (components/results): 1 device: pusher / dust or dirt problem identified.1 device: wiring harness / device migration.There was no remedial action taken.This device is not labeled for single use.
|
| |
|
Event or Problem Description
|
|
This record is a consolidation of records summarized as a part of the fda voluntary malfunction summary reporting program.Events occurred between july 1-september 30, 2025.This report summarizes 2 malfunction event(s), where it was reported the device experienced unintended direction of the zoom; unintended motion.No adverse consequence or clinically relevant delay in treatment was reported.
|
| |
|
Search Alerts/Recalls
|