| Model Number |
PM2160 |
| Medical Device Problem Codes |
Premature Discharge of Battery (1057); Failure to Interrogate (1332); Unexpected Mode Switch (4076)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that the patient presented for remote follow up via merlin.Net.A review of the transmission revealed that the implantable cardioverter defibrillator (icd) was in backup operation.The patient subsequently reported in-clinic and there was no magnet response observed from the device.Premature battery depletion was suspected.The device was explanted and replaced.The patient was stable.
|
| |
|
Additional Manufacturer Narrative
|
|
B5 updated to state device in question was a pacemaker and not an implantable cardioverter defibrillator.
|
| |
|
Event or Problem Description
|
|
It was reported that the patient presented for remote follow up via merlin.Net.A review of the transmission revealed that the pacemaker (pm) was in backup operation.The patient subsequently reported in-clinic and there was no magnet response or inductive telemetry observed from the device.Premature battery depletion was suspected.The device was explanted and replaced.The patient was stable.
|
| |
|
Additional Manufacturer Narrative
|
|
The reported events of backup operation and unable to interrogate was confirmed.As received, the device had no telemetry communication and no output.Visual inspection of the header attachment area detected a bonding anomaly.A device history record (dhr) review was performed and all required manufacturing processes and inspections steps were confirmed to be completed per the requirements.A device hermeticity breach was observed, consistent with feedthrough damage as a result of fluid intrusion between the header and case, and subsequent fluid ingress to the internal electronics.The device was cut open to enable further testing and battery was found depleted.Hybrid circuitry was tested, resulting in high current drain, consistent with moisture damage, depleting the battery and resulting in the reported events.A manufacturing anomaly may have occurred, which resulted in the header bonding anomaly.The device is included in the assurity and endurity pacemakers' header anomaly advisory issued by abbott on 15 march 2021.
|
| |
|