• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BIOTRONIK SE & CO. KG BIOMONITOR III; IMPLANTABLE CARDIAC MONITOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BIOTRONIK SE & CO. KG BIOMONITOR III; IMPLANTABLE CARDIAC MONITOR Back to Search Results
Model Number 436066
Medical Device Problem Code Premature Elective Replacement Indicator (1483)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/15/2025
Type of Reportable Event Malfunction
Event or Problem Description
This device was explanted due to eri with unexpected battery behavior.No adverse patient events were reported.Should additional information become available, this file will be updated.
 
Additional Manufacturer Narrative
The device was not returned for analysis.The analysis is thus based on the inspection of the quality documents associated with the manufacture of this particular device as well as the data returned for evaluation.The manufacturing process for this device was reviewed and all production steps were performed accordingly.There was no sign of any inconsistency during the manufacturing process.Particularly the final acceptance test proved the device functions to be as specified.The analysis revealed an unexpected battery behavior, which led to the clinical observation.Based on the data available for analysis the root cause for that behavior was not determinable and can only be clarified by an analysis of the device itself.However, it cannot be excluded that this occurrence resulted from a defective component.Should the device become available for analysis, this investigation will be updated.
 
Additional Manufacturer Narrative
The device was received for analysis.Prior to the analysis of the device, the quality documents accompanying the manufacturing process for this device were reviewed and all production steps were performed accordingly.In particular, the final acceptance test proved the device functions to be as specified.During initial device interrogation the battery status was eri, which was detected on february 28, 2025.The memory content documented a normal and expected current consumption, however, an inconsistency between current consumption and battery voltage was noted.Therefore the device was opened and subjected to further investigation.The visual inspection of the inner assembly showed no anomalies.The measurement of the battery voltage showed a depleted battery.In a next step, the battery was disconnected from the electronic module.Further thorough investigation of the electronic module did not show any anomalies.In particular the current consumption proved to be normal and expected.The battery was sent to the manufacturer for further analysis.The manufacturing records of the battery were inspected, documenting that the battery parameters were within specification during the manufacturing.The battery was subjected to an electrical inspection as well as a microcalorimetry analysis.The microcalorimetry analysis showed a normal heat dissipation.In a next step, the battery was opened for destructive analysis.Inspection of the inner assembly revealed evidence of an internal short circuit, which led to an elevated internal self-depletion within the battery and therefore contributed to the clinical observation.In conclusion, an elevated internal self-depletion within the battery was found to be the root cause for the clinical observation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BIOMONITOR III
Common Device Name
IMPLANTABLE CARDIAC MONITOR
Manufacturer (Section D)
BIOTRONIK SE & CO. KG
woermannkehre 1
berlin 12359
Manufacturer Contact
6024 jean road
lake oswego, OR 97035
8772459800
MDR Report Key23397772
Report Number1028232-2025-05249
Device Sequence Number15956264
Product Code MXD
UDI-Device Identifier04035479158453
UDI-Public(01)04035479158453(17)240131
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
K190548
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Device Explanted Year2025
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 03/19/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date01/31/2024
Device Model Number436066
Device Catalogue Number436066
Was Device Available for Evaluation? Device Returned to Manufacturer
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/24/2025
Supplement Date Received by Manufacturer10/30/2025
03/19/2026
Initial Report FDA Received Date10/27/2025
Supplement Report FDA Received Date10/30/2025
03/19/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured07/08/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization;
Patient Age82 YR
Patient SexFemale
-
-