The device was not returned for analysis.The analysis is thus based on the inspection of the quality documents associated with the manufacture of this particular device as well as the data returned for evaluation.The manufacturing process for this device was reviewed and all production steps were performed accordingly.There was no sign of any inconsistency during the manufacturing process.Particularly the final acceptance test proved the device functions to be as specified.The analysis revealed an unexpected battery behavior, which led to the clinical observation.Based on the data available for analysis the root cause for that behavior was not determinable and can only be clarified by an analysis of the device itself.However, it cannot be excluded that this occurrence resulted from a defective component.Should the device become available for analysis, this investigation will be updated.
|
|
The device was received for analysis.Prior to the analysis of the device, the quality documents accompanying the manufacturing process for this device were reviewed and all production steps were performed accordingly.In particular, the final acceptance test proved the device functions to be as specified.During initial device interrogation the battery status was eri, which was detected on february 28, 2025.The memory content documented a normal and expected current consumption, however, an inconsistency between current consumption and battery voltage was noted.Therefore the device was opened and subjected to further investigation.The visual inspection of the inner assembly showed no anomalies.The measurement of the battery voltage showed a depleted battery.In a next step, the battery was disconnected from the electronic module.Further thorough investigation of the electronic module did not show any anomalies.In particular the current consumption proved to be normal and expected.The battery was sent to the manufacturer for further analysis.The manufacturing records of the battery were inspected, documenting that the battery parameters were within specification during the manufacturing.The battery was subjected to an electrical inspection as well as a microcalorimetry analysis.The microcalorimetry analysis showed a normal heat dissipation.In a next step, the battery was opened for destructive analysis.Inspection of the inner assembly revealed evidence of an internal short circuit, which led to an elevated internal self-depletion within the battery and therefore contributed to the clinical observation.In conclusion, an elevated internal self-depletion within the battery was found to be the root cause for the clinical observation.
|