Model Number 1C8363
Medical Device Problem Code Material Split, Cut or Torn (4008)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that fourteen (14) extension sets had slits on the filter.This was discovered during setup/preparation.There was no patient involvement.No additional information is available.
 
Additional Manufacturer Narrative
Should additional relevant information become available, a supplemental report will be submitted.
 
Additional Manufacturer Narrative
Additional information: d4: expiration date (update the null value to n/a), h4, h6 (update codes) and h11.H11: the actual devices, a video and photographs were received for evaluation.The visual inspection of the photos revealed marks at the junction of both ends of the filter.According to the component specifications, these marks are not considered defects.However, the visual inspection of the actual samples did not identify any abnormalities that could have contributed to the reported condition.Pressure testing was performed on the samples, and no leaks were observed.The reported condition was not verified.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NA
Common Device Name
SET, ADMINISTRATION, INTRAVASCULAR
Manufacturer (Section D)
BAXTER HEALTHCARE CORPORATION
deerfield IL
Manufacturer (Section G)
BAXTER HEALTHCARE - HAINA
parque industrial itabo
carretera sanchez km 18 1/2
haina 21426
DR   21426
Manufacturer Contact
25212 w. illinois route 120
round lake, IL 60073
2242702068
MDR Report Key23398255
Report Number1416980-2025-05298
Device Sequence Number18645713
Product Code FPA
UDI-Device Identifier00085412013718
UDI-Public(01)00085412013718(10)DR24K11055(17)291130
Combination Product (Y/N)N
Initial Reporter CountryHK
PMA/510(K) Number
K153158
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date (Section B) 11/20/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Model Number1C8363
Device Catalogue Number1C8363
Device Lot NumberDR24K11055
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/29/2025
Supplement Date Received by Manufacturer11/20/2025
Initial Report FDA Received Date10/27/2025
Supplement Report FDA Received Date11/20/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured11/15/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-