• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMVIE US CORP LLC IMP, TSV, 3.7, 13, MTXF, MG,HA; DENTAL IMPLANT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMVIE US CORP LLC IMP, TSV, 3.7, 13, MTXF, MG,HA; DENTAL IMPLANT Back to Search Results
Catalog Number TSVTH13
Medical Device Problem Code Mechanical Problem (1384)
Health Effect - Clinical Code Pain (1994)
Date of Event 03/26/2025
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Zimvie complaint number (b)(4).A1: patient identifier unknown / not provided.A2: age at time of the event unknown / not provided.D10.Concomitant medical products tsvth13, imp, tsv, 3.7, 13, mtxf, mg, ha lot 1265965.
 
Event or Problem Description
The doctor reports that one of the two implants presents damaged threads and the other implant had to be removed due to an infection.The doctor reports pain.The procedure was not completed.The affected dental positions are 13 and 26.This complaint refers to the implant which was removed due to damage threads.
 
Additional Manufacturer Narrative
Zimvie complaint number (b)(4).The following fields have been updated: b4: date of this report.B5: describe event or problem.D3.Manufacturer email address.G1.Contact office name and email address.G1.Contact office - manufacturer site - email address.G3: date received by manufacturer.G6: type of report.H1: type of reportable event.H2: follow up type.H3: device evaluated by manufacturer.H6: adverse event problem.H10: additional narrative.Zimvie received one (1) tsvth13, (imp, tsv, 3.7, 13, mtxf, mg, ha) for evaluation.Visual evaluation was performed, the implant had signs of use with debris on its external threads.The implants drive feature was damaged.Dhr review was completed for the subject lot number 1265965.No deviations or non-conformances, which could have caused or contributed to the reported event, were noted as part of the dhr.Complaint history review was performed for the reported lot number 1265965 for similar events and no other complaint was identified.Review completed utilizing keywords: ¿dental: functional: damaged drive feature.¿ based on the investigation and risk management file review, the most likely root cause determined from the investigation was customer error, applying to much torque/speed during seating.Therefore, based on the available information, a device malfunction did occur.The implants drive feature was damaged.The reported event was confirmed.No further investigation or immediate capa / hhe/d escalation is required, as the complaint investigation did not confirm the products were nonconforming at the time of distribution, and no new failure mode, harm, or hazardous situation was identified through the investigation performed.At this time, the complaint investigation has been completed and the record will be closed.If additional information is received, the record will be re-opened for further evaluation.
 
Event or Problem Description
No further event information is available at the time of this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
IMP, TSV, 3.7, 13, MTXF, MG,HA
Common Device Name
DENTAL IMPLANT
Manufacturer (Section D)
ZIMVIE US CORP LLC
4555 riverside drive
palm beach gardens FL 33410
Manufacturer (Section G)
ZIMVIE US CORP LLC
4555 riverside drive
palm beach gardens FL 33410
Manufacturer Contact
mark mashburn
4555 riverside drive
palm beach gardens, FL 33410
5619713230
MDR Report Key23400067
Report Number0001038806-2025-02723
Device Sequence Number10015906
Product Code DZE
UDI-Device Identifier00889024344310
UDI-Public(01)00889024344310(17)280513(10)1265965
Combination Product (Y/N)N
Initial Reporter CountryIT
PMA/510(K) Number
K133339
Exemption Number5645646
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2025
Device Explanted Year2025
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Distributor
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 11/27/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Other
Device Catalogue NumberTSVTH13
Device Lot Number1265965
Was Device Available for Evaluation? No
Date Returned to Manufacturer10/01/2025
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer Not provided
Supplement Date Received by ManufacturerNot provided
Initial Report FDA Received Date10/28/2025
Supplement Report FDA Received Date11/27/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured05/15/2023
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
DENTAL IMPLANT SEE H10 NARRATIVE.
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
Patient EthnicityNon Hispanic
-
-