W. L. GORE & ASSOCIATES, INC. GORE® EXCLUDER® AAA ENDOPROSTHESIS; SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT
|
Back to Search Results |
|
| Catalog Number |
PLC201000 |
| Medical Device Problem Codes |
Break (1069); Activation Failure (3270)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/03/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
On (b)(6) 2025, the patient underwent an endovascular procedure to treat an abdominal aortic aneurysm (aaa) utilizing gore® excluder® aaa endoprosthesis (contralateral leg).Reportedly, after the contralateral leg was inserted, the deployment string broke at the distal area of the graft, thus, the device did not deploy at all.The contralateral leg was removed, discarded, and replaced with another contralateral leg.The cause of the deployment string breakage is unknown and the anatomy was a little tortuous but not bad.The patient tolerated the procedure.
|
| |
|
Additional Manufacturer Narrative
|
|
H6: code c19 - a review of the manufacturing records indicated the lot met all pre-release specifications.H6: code c20 -the device was discarded at the treating facility and was therefore not available for engineering evaluation by gore.H6: code d15 - there is insufficient information available for gore to reasonably draw conclusions related to aspects of the event, therefore conclusion code ¿d15: cause not established¿ is being used.Insufficient information may include limited or missing relevant medical records, involvement of multiple implanted devices (including non-gore devices) in the field of treatment, patient non-compliance, and/or a general lack of available detail or specificity related to an adverse event and/or device.W.L.Gore & associates, inc.(gore) is submitting this report to comply with 21 c.F.R.Part 803, the medical device reporting regulation.This report is based upon information obtained by gore, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Blank fields present on this report include required fields and fields determined to be not applicable.Blank required fields indicate that the information was not provided, was deemed unavailable or was not applicable.This report does not constitute an admission or a conclusion by fda, gore, or its associates that the device, gore or its associates caused or contributed to the event described in the report.In particular, this report does not constitute a legal admission by anyone that the product described in this report has any defects or has malfunctioned, as defined from a legal standpoint.These words are included in the report and are fixed items for selection created by the fda, to categorize the type of event solely for the purpose of reporting pursuant to part 803.This statement should be included with any information or report disclosed to the public under the freedom of information act.
|
| |
|
Search Alerts/Recalls
|
|
|