AIZU OLYMPUS CO., LTD. EVIS EXERA II COLONOVIDEOSCOPE Back to Search Results
Model Number CF-H180AI
Medical Device Problem Code Microbial Contamination of Device (2303)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/18/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The investigation is ongoing.A supplemental report will be submitted when the investigation is completed or if additional information becomes available.
 
Event or Problem Description
It was reported that during reprocessing, the colonovideoscope tested positive for 8 colony forming units (cfus) of enterobacteriaceae.All channels were sampled.The user did not report any contamination or any other serious deterioration in state of health of any person, to which the scope could have been a contributory cause.
 
Additional Manufacturer Narrative
B5: no additional information received from customer.D8: additional information d9: additional information h2: additional information, device evaluation h3: additional information h6: additional information this report is being supplemented to provide additional information based on the results of third-party testing, the device evaluation and the complaint investigation.The device was evaluated where no abnormalities were found that could have led to the reported event.Based on the results of the investigation, the reported event could not be confirmed, and a root cause could not be determined.Growth of microorganisms were reported through culture testing by the user after reprocessing.However, when olympus culture tested after reprocessing in accordance with the instructions for use (ifu) before repair, the results conformed to the regulation's recommendation.Olympus will continue to monitor field performance for this device.
 
Event or Problem Description
No additional information received from customer.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EVIS EXERA II COLONOVIDEOSCOPE
Common Device Name
COLONOVIDEOSCOPE
Manufacturer (Section D)
AIZU OLYMPUS CO., LTD.
3-1-1 niiderakita
aizuwakamatsu-shi, fukushima 965-8 520
JA  965-8520
Manufacturer (Section G)
AIZU OLYMPUS CO., LTD.
3-1-1 niiderakita
aizuwakamatsu-shi, fukushima 965-8 520
JA   965-8520
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key23403030
Report Number9610595-2025-28086
Device Sequence Number21344052
Product Code FDF
UDI-Device Identifier04953170339585
UDI-Public04953170339585
Combination Product (Y/N)N
Initial Reporter CountryFR
PMA/510(K) Number
K100584
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility
Initial Reporter Occupation Administrator/Supervisor
Type of Report Initial,Followup
Report Date (Section B) 04/17/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberCF-H180AI
Was Device Available for Evaluation? Device Returned to Manufacturer
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 09/30/2025
Supplement Date Received by Manufacturer03/23/2026
Initial Report FDA Received Date10/28/2025
Supplement Report FDA Received Date04/17/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured09/27/2016
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-