|
The lens was returned for evaluation.Solution is dried on the lens.Both haptics are broken in the gusset area.Only one haptic was returned and it was adhered to the optic by solution.The optic edge is broken in two areas with loose pieces of lens.A crack is observed near one of the broken areas.A device history record review and a non-conformance review of the reported lot number was conducted.No deviations were identified and all corresponding production release specifications defined in the device master record were met.Qualified associated products were indicated.Broken haptic damage was observed.All lenses are 100% inspected for cosmetic attributes and the damage exhibited by the returned complaint sample would not have met alcon¿s release criteria.Based on our observation, it can be reasonably concluded that the damage is not manufacturing related.Due to the presence of surgical solution and the condition of the returned sample, the damage is most likely related to customer handling.Qualified associated products were indicated.The ifu instructs: using holding forceps, grasp the lens by the optic edge and gently place the lens anterior side up into the back of the ovd-filled cartridge.The lens should be inserted until the optic is a little more than half-way inside the cartridge.Use the holding forceps to gently push down on the lens, verifying that the lens is on the bottom surface of the cartridge.Using holding forceps, take the trailing haptic, and gently fold the haptic onto the anterior side of the optic.Slowly grip or push the optic edge to position the lens as far into the cartridge as the forceps will permit, while ensuring the lens remains on the bottom surface of the cartridge and the trailing haptic remains on the optic.Failure to follow these steps may cause the lens to advance incorrectly causing delivery issues and/or damage.Company lens ifu: follow the section regarding directions for use for information on the maximum allowed time for the iol to stay in the folded condition.Failure to adhere to manufacturer¿s recommendations may result in iol damage.Ifu note: during lens loading and insertion, do not allow the company iol to remain in a folded condition within the selected iol delivery system for more than 3 minutes prior to completing insertion into the capsular bag.The manufacturer internal reference number is: (b)(4).
|