• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TORNIER S.A.S. TORNIER PERFORM ANATOMIC GLENOID PEGGED S40; PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TORNIER S.A.S. TORNIER PERFORM ANATOMIC GLENOID PEGGED S40; PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED Back to Search Results
Catalog Number DWE303
Medical Device Problem Code Loss of Osseointegration (2408)
Health Effect - Clinical Codes Pain (1994); Inadequate Osseointegration (2646)
Date of Event 09/03/2025
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that, the glenoid implant was dislocated.The revision surgery involved removing the implant and performing an iliac bone graft.
 
Additional Manufacturer Narrative
Additional information has been requested and if received, will be provided in the supplemental report.
 
Additional Manufacturer Narrative
The reported event could not be confirmed since the device was not returned for evaluation and no other evidences were provided.A device inspection was not possible since the affected device was not returned.More detailed information about the complaint event as well as the affected device must be available in order to determine the root cause of the complaint event.A review of the device history for the reported lot did not indicate any abnormalities.No corrective actions are required at this time.A review of the labeling did not indicate any abnormalities.No indications of material, manufacturing or design related problems were found during the investigation.If the device is returned or if any additional information is provided, the investigation will be reassessed.
 
Event or Problem Description
It was reported that, the glenoid implant was dislocated.The revision surgery involved removing the implant and performing an iliac bone graft.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TORNIER PERFORM ANATOMIC GLENOID PEGGED S40
Common Device Name
PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED
Manufacturer (Section D)
TORNIER S.A.S.
161 rue lavoisier
montbonnot saint-martin 38330
FR  38330
Manufacturer (Section G)
WRIGHT MEDICAL CORK (TORNIER ORTHOPEDICS IRELAND LTD)
harnetts cross
macroom, co. cork NA
EI   NA
Manufacturer Contact
dannielle stap
325 corporate drive
mahwah, NJ 07430
2018315000
MDR Report Key23405624
Report Number3000931034-2025-00587
Device Sequence Number14678048
Product Code KWS
UDI-Device Identifier03700386937354
UDI-Public(01)03700386937354(17)260708(21)AD6986017
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K111902
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Device Explanted Year2025
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 12/23/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date07/08/2026
Device Catalogue NumberDWE303
Device Lot NumberAD6986
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 09/29/2025
Supplement Date Received by Manufacturer12/04/2025
Initial Report FDA Received Date10/28/2025
Supplement Report FDA Received Date12/23/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured07/08/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age55 YR
Patient SexFemale
Patient Weight53 KG
-
-