| Catalog Number |
656322S |
| Medical Device Problem Code |
Positioning Failure (1158)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/29/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Additional information has been requested and if received, will be provided in the supplemental report.
|
| |
|
Event or Problem Description
|
|
As reported: "variax 2 dr case during distal screw insertion, a 22mm locking screw was inserted into the second hole from the radial side.However, it failed to lock.An attempt was made to remove it, but the screw spun freely within the screw hole and could not be removed.As it could not be removed, the surgery was completed while the unlocked screw was left in place.".
|
| |
|
Additional Manufacturer Narrative
|
|
The reported event could not be confirmed since the device was not returned for evaluation and no other evidences were provided.More detailed information about the complaint event as well as the affected device must be available in order to determine the root cause of the complaint event.A review of the device history for the reported lot did not indicate any abnormalities.No corrective actions are required at this time.A review of the labeling did not indicate any abnormalities.No indications of material, manufacturing or design related problems were found during the investigation.If the device is returned or if any additional information is provided, the investigation will be reassessed.
|
| |
|
Event or Problem Description
|
|
As reported: "variax 2 dr case during distal screw insertion, a 22mm locking screw was inserted into the second hole from the radial side.However, it failed to lock.An attempt was made to remove it, but the screw spun freely within the screw hole and could not be removed.As it could not be removed, the surgery was completed while the unlocked screw was left in place.".
|
| |
|