• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: PHILIPS ULTRASOUND, INC CX50 DIAGNOSTIC ULTRASOUND SYSTEM; SYSTEM, IMAGING, PULSED DOPPLER, ULTRASONIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PHILIPS ULTRASOUND, INC CX50 DIAGNOSTIC ULTRASOUND SYSTEM; SYSTEM, IMAGING, PULSED DOPPLER, ULTRASONIC Back to Search Results
Model Number 795087
Medical Device Problem Code Application Program Freezes or Fails to Launch (4031)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/29/2025
Type of Reportable Event Malfunction
Event or Problem Description
It was reported the cx50 ultrasound system was not available for an emergency surgery.It was discovered that the ultrasound system was unable to turn on during set-up.A philips field service engineer secured the power supply connector to resolve the customer¿s immediate concerns.There was no reported patient nor user harm as a result of the issue.Philips has started an investigation, and upon completion, a follow up report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CX50 DIAGNOSTIC ULTRASOUND SYSTEM
Common Device Name
SYSTEM, IMAGING, PULSED DOPPLER, ULTRASONIC
Manufacturer (Section D)
PHILIPS ULTRASOUND, INC
22100 bothell everett highway
bothell WA 98021
Manufacturer (Section G)
PHILIPS ULTRASOUND, INC
22100 bothell everett highway
bothell WA 98021
Manufacturer Contact
roxanne ramirez
22100 bothell everett highway
bothell, WA 98021
MDR Report Key23408135
Report Number3019216-2025-000398
Device Sequence Number21098908
Product Code IYN
UDI-Device Identifier00884838097964
UDI-Public(01)00884838097964
Combination Product (Y/N)N
Initial Reporter CountryUK
PMA/510(K) Number
K123754
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 10/28/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number795087
Device Catalogue Number795087
Was Device Available for Evaluation? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/29/2025
Initial Report FDA Received Date10/28/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured11/03/2022
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-