DEPUY ORTHOPAEDICS INC US APEX HOLE ELIM POSITIVE STOP; HIP ACETABULAR CUP Back to Search Results
Catalog Number 124603000
Medical Device Problem Codes Device Dislodged or Dislocated (2923); Adverse Event Without Identified Device or Use Problem (2993); Noise, Audible (3273)
Health Effect - Clinical Code Pain (1994)
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Product complaint #(b)(4).Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees, caused or contributed to the potential event described in this report.H11 additional narrative: d1, d2, d3, d4, g4-510k: this report is for an unk hip acetabular cup/unknown lot.Part and lot number are unknown.Without the specific part number, the udi number and 510-k number is unknown.
 
Event or Problem Description
It was reported that on (b)(6) 2014, the patient underwent a tha surgery for oa with the implants in question.After the surgery in (b)(6) 2024, x-rays showed no abnormalities in the liner in question.In (b)(6) 2025, ct scan revealed some wear (eccentricity of the head) movement behind the liner.On (b)(6) 2025, the patient visited the hospital complaining of abnormal noise and pain.An x-ray revealed significant displacement of the femoral head center.Dislocation and wear were suspected, but the x-ray findings did not reveal dislocation.The surgeon doubted that such extensive wear could occur in such a short period of time, and suspected liner dislocation.On (b)(6), a revision was scheduled for (b)(6) 2025.During the surgery, a sales representative checked the situation from the surgical field and confirmed the following: the liner at the top of the cup rim was embedded into the cup, with the anti-rotation tab destroyed.The delta head had collided with the cup rim and the tapered locking portion of the cup/liner (discolored due to friction with the head), causing significant damage.The surgeon determined that re-fixing the liners would be difficult, so all implants placed in the previous operation were removed and a complete revision was performed.The surgery was completed successfully with 30 minutes surgical delay.No further information is available.
 
Additional Manufacturer Narrative
Product complaint #(b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information that has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint recon described in this report.H11 additional narrative: added: d4 (lot catalog), h4, h6 medical device problem code.Corrected: d1, d2a, d4 primary udi number.If information that was not available for the initial report is obtained, a follow-up report will be filed as appropriate.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information that has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint recon described in this report.H11 additional narrative: added: d9.If information that was not available for the initial report is obtained, a follow-up report will be filed as appropriate.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Investigation summary: it was reported that on (b)(6) 2014, the patient underwent a tha surgery for oa with the implants in question.After the surgery in (b)(6) 2024, x-rays showed no abnormalities in the liner in question.In (b)(6) 2025, ct scan revealed some wear (eccentricity of the head) movement behind the liner.On (b)(6) 2025, the patient visited the hospital complaining of abnormal noise and pain.An x-ray revealed significant displacement of the femoral head center.Dislocation and wear were suspected, but the x-ray findings did not reveal dislocation.The surgeon doubted that such extensive wear could occur in such a short period of time, and suspected liner dislocation.On october 6th, a revision was scheduled for (b)(6) 2025.During the surgery, a sales representative checked the situation from the surgical field and confirmed the following: the liner at the top of the cup rim was embedded into the cup, with the anti-rotation tab destroyed.·the delta head had collided with the cup rim and the tapered locking portion of the cup/liner (discolored due to friction with the head), causing significant damage.The surgeon determined that re-fixing the liners would be difficult, so all implants placed in the previous operation were removed and a complete revision was performed.The surgery was completed successfully with 30 minutes surgical delay.The product was returned to depuy synthes for evaluation.Visual inspection of the returned device found nothing indicative of a device nonconformance.A functional test was not performed since it was not applicable to the complaint condition.A dimensional inspection was not performed since it was not applicable to the complaint condition.The overall complaint was not confirmed as the observed condition of the apex hole elim positive stop would have not contributed to the reported adverse event.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.There is no indication that a design or manufacturing issue has caused the reported complaint condition.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device history lot: a manufacturing record evaluation was performed for the finished device lot number d13113401, and no non-conformances / manufacturing irregularities were identified.Added: d10 corrected: d4 (primary udi #) h3.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
APEX HOLE ELIM POSITIVE STOP
Common Device Name
HIP ACETABULAR CUP
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
Manufacturer (Section G)
JTE WARSAW MFG SITE
700 orthopaedic drive
warsaw IN 46582
Manufacturer Contact
elba bello
700 orthopaedic dr.
warsaw, IN 46581
9083863534
MDR Report Key23408526
Report Number1818910-2025-18500
Device Sequence Number9076080
Product Code LPH
UDI-Device Identifier10603295019688
UDI-Public(01)10603295019688(17)231031(10)D13113401
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K963309
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2014
Device Explanted Year2025
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date10/31/2023
Device Catalogue Number124603000
Device Lot NumberD13113401
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 10/09/2025
Supplement Date Received by Manufacturer10/29/2025
11/25/2025
01/27/2026
Initial Report FDA Received Date10/28/2025
Supplement Report FDA Received Date11/03/2025
11/26/2025
01/28/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured04/20/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
APEX HOLE ELIM POSITIVE STOP; C-STEM SZ3; DELTA CER HEAD 9/10 28MM -3; PINN CAN BONE SCREW 6.5MMX20MM; PINN CAN BONE SCREW 6.5MMX20MM; PINN MAR ETO NEUT 28IDX48OD; PINNACLE SECTOR II CUP 48MM
Outcome Attributed to Adverse Event Required Intervention;
Patient Age83 YR
Patient SexFemale
-
-