• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: NIPRO (THAILAND) CORP., LTD NIPRO SURDIAL DX HEMODIALYSIS SYSTEM; HYPODERMIC NEEDLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

NIPRO (THAILAND) CORP., LTD NIPRO SURDIAL DX HEMODIALYSIS SYSTEM; HYPODERMIC NEEDLE Back to Search Results
Model Number MC+SDX01
Medical Device Problem Code Detachment of Device or Device Component (2907)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/08/2025
Type of Reportable Event Malfunction
Event or Problem Description
Reported condition: front rotor panel flat head screw is missing.The national service engineer installed new blood pump rotor assembly serial#: (b)(6) and perform blood pump occlusion and adjustment procedure.Machine operates per manufacturer specifications.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NIPRO SURDIAL DX HEMODIALYSIS SYSTEM
Common Device Name
HYPODERMIC NEEDLE
Manufacturer (Section D)
NIPRO (THAILAND) CORP., LTD
10/2 moo 8
bangnomko
sena, ayuthaya 13110
TH  13110
MDR Report Key23410587
Report Number1056186-2025-00037
Device Sequence Number8983363
Product Code KDI
UDI-Device Identifier00383790008815
UDI-Public00383790008815
Combination Product (Y/N)N
Initial Reporter StateOH
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Importer
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 10/28/2025
Report Date (Section F) 10/28/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberMC+SDX01
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? Yes
Distributor Facility Aware Date10/08/2025
Event Location Outpatient Treatment Facility
Date Report to Manufacturer10/28/2025
Initial Date Received by Manufacturer Not provided
Initial Report FDA Received Date10/28/2025
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
Date Report Sent to FDA10/28/2025
-
-