Model Number MC+SDX01
Medical Device Problem Code Detachment of Device or Device Component (2907)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/08/2025
Type of Reportable Event Malfunction
Event or Problem Description
Reported condition: front rotor panel flat head screw is missing.The national service engineer installed new blood pump rotor assembly serial# (b)(6) and perform blood pump occlusion and adjustment procedure.Machine operates per manufacturer specifications.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NIPRO SURDIAL DX HEMODIALYSIS SYSTEM
Common Device Name
HYPODERMIC NEEDLE
Manufacturer (Section D)
NIPRO (THAILAND) CORP., LTD
10/2 moo 8
bangnomko
sena, ayuthaya 13110
TH  13110
Manufacturer (Section G)
SHIBUYA (SBY)
2-72, wakamiya
kanazawa, isikawa 920-0 054
JA   920-0054
Manufacturer Contact
michelle tejada
3150 nw 107th avenue
miami, FL 33172
3055997174
MDR Report Key23411278
Report Number3016250252-2025-00016
Device Sequence Number18413262
Product Code KDI
UDI-Device Identifier00383790008815
UDI-Public00383790008815
Combination Product (Y/N)N
Initial Reporter StateOH
Initial Reporter CountryUS
PMA/510(K) Number
K182940
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 10/28/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberMC+SDX01
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer Not provided
Initial Report FDA Received Date10/29/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-