Model Number 6893200
Medical Device Problem Code Failure of Device to Self-Test (2937)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/02/2025
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the ventilator failed flow transducer test during pre-use check.There was no patient involvement.Manufacturer's ref.#: (b)(4).
 
Event or Problem Description
Manufacturer's ref.#: (b)(4).
 
Additional Manufacturer Narrative
The field service engineer (fse) has investigated the reported ventilator on-site.The fse replaced the o2 gas module to resolve the issue and sent back for further investigation.After the replacement, the ventilator passed all functional and safety tests and was cleared for clinical use.The provided logs confirm the reported issue.The returned o2 gas module has been tested in a ventilator.It passed the pre-use check.No issues were found during ventilation for 4 days.It passed another pre-use check after ventilation without any issue.The air and o2 gas modules regulate the inspiratory gas flow and gas mixture.The faulty gas module may lead to stop of ventilation, low o2 or high pressure.The conclusion is that the reported ventilator has failed to meet its specifications during the reported event.Even though, the reported issue could not be reproduced at the manufacturer site, the replaced part still could be the cause of the issue as we know this type of failure is an intermittent failure.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SERVO-AIR LITE
Common Device Name
VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT, FACILITY USE
Manufacturer (Section D)
MAQUET CRITICAL CARE AB
roentgenvagen 2
solna
Manufacturer (Section G)
MAQUET CRITICAL CARE AB
roentgenvagen 2
solna
Manufacturer Contact
caroline kabbabe
roentgenvagen 2
solna 
MDR Report Key23411562
Report Number8010042-2025-0001590
Device Sequence Number10407719
Product Code MNT
UDI-Device Identifier07325710015117
UDI-Public(01)07325710015117(11)250311
Combination Product (Y/N)N
Initial Reporter StateNC
Initial Reporter CountryUS
PMA/510(K) Number
K230173
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source User Facility,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 10/29/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number6893200
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/02/2025
Supplement Date Received by Manufacturer02/13/2026
Initial Report FDA Received Date10/29/2025
Supplement Report FDA Received Date02/13/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured03/11/2025
Is the Device Labeled for Single Use? (Y/N) No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-