| Model Number |
24967 |
| Medical Device Problem Codes |
Communication or Transmission Problem (2896); Output Problem (3005)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/17/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that during a implantation, the patient connector lost connection to the mobile programmer application and the implantable device.It was noted that the telemetry strength fluctuated from three bars to one bar, then lost connection entirely without any apparent movement or change.It was further noted that the issue was particularly evident during threshold testing with drop before the threshold test could be completed and the electrograms (egm) display intermittently dropped out, showing white dots.Troubleshooting steps were taken to include swapping out the application, however connection continued to drop out.No patient complications have been reported as a result of this event.Application version: 4.4.2.
|
| |
|
Additional Manufacturer Narrative
|
|
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Additional Manufacturer Narrative
|
|
Product analysis: performance data collected from the mobile programmer was received and analyzed.Analysis of the service logs found the customer comment that the patient connector lost connection was confirmed.Analysis of the service logs found the customer comment that the electrograms (egm) display intermittently dropped out was confirmed.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Search Alerts/Recalls
|