• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY IRELAND - 9616671 ATTUNE PS FEM RT SZ 6 NAR CEM; ATTUNE IMPLANT : KNEE FEMORAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY IRELAND - 9616671 ATTUNE PS FEM RT SZ 6 NAR CEM; ATTUNE IMPLANT : KNEE FEMORAL Back to Search Results
Catalog Number 150410226
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Pain (1994); Loss of Range of Motion (2032); Joint Laxity (4526)
Date of Event 01/01/2025
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that on an unknown date, the patient experienced excessive laxity in extension and deficit in flexion on a tka with attune from 2022.Patient experienced reduced mobility and pain and required revision surgery.
 
Additional Manufacturer Narrative
Product complaint #: (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.H11 additional narrative: d4: udi: as the lot number for the device involved in the event was not provided, the full udi is currently not available.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Investigation summary- no device associated with this report was received for examination.An evaluation of the manufacturing record could not be performed as the required product/lot number was not provided to complete the evaluation.As part of our company quality system process, all devices are manufactured, inspected, and distributed to approved specifications.Additional complaint information monitoring for potential safety signals will be conducted through complaint trending as part of the post-market surveillance.If additional information is made available, the investigation will be updated as applicable.Device history lot- the device lot number is unknown, therefore a¿device history review could not be performed.If the lot/serial number becomes available, the record will be re-assessed.H11 additional narrative: added: h6 health effect - clinical code.
 
Additional Manufacturer Narrative
Product complaint #
=
>(b)(4).Investigation summary
=
> no device associated with this report was received for examination.An evaluation of the manufacturing record could not be performed as the required product/lot number was not provided to complete the evaluation.As part of our company quality system process, all devices are manufactured, inspected, and distributed to approved specifications.Additional complaint information monitoring for potential safety signals will be conducted through complaint trending as part of the post-market surveillance.If additional information is made available, the investigation will be updated as applicable.Device history lot
=
> the device lot number is unknown, therefore a device history review could not be performed.If the lot/serial number becomes available, the record will be re-assessed.H11 additional narrative: added: h6 health effect clinical code.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ATTUNE PS FEM RT SZ 6 NAR CEM
Common Device Name
ATTUNE IMPLANT : KNEE FEMORAL
Manufacturer (Section D)
DEPUY IRELAND - 9616671
loughbeg ringaskiddy co.
cork
EI 
Manufacturer (Section G)
JTE WARSAW MFG SITE
700 orthopaedic drive
warsaw IN 46582
Manufacturer Contact
elba bello
700 orthopaedic dr.
warsaw, IN 46581
9083863534
MDR Report Key23413376
Report Number1818910-2025-18531
Device Sequence Number9084581
Product Code JWH
UDI-Device Identifier10603295041856
UDI-Public(01)10603295041856
Combination Product (Y/N)N
Initial Reporter CountrySZ
PMA/510(K) Number
P830055
Number of Events Summarized1
Summary Report (Y/N)N
Device Explanted Year2025
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Catalogue Number150410226
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 10/10/2025
Supplement Date Received by Manufacturer11/05/2025
12/02/2025
Initial Report FDA Received Date10/29/2025
Supplement Report FDA Received Date11/06/2025
12/04/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-