• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEROYAL CIENTIFICA DE LATINOAMERICA S.R.L SKIN TEMPERATURE PROBE, 400 SERIES; CONTINUOUS MEASUREMENT THERMOMETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEROYAL CIENTIFICA DE LATINOAMERICA S.R.L SKIN TEMPERATURE PROBE, 400 SERIES; CONTINUOUS MEASUREMENT THERMOMETER Back to Search Results
Model Number 81-010400
Medical Device Problem Code Incorrect Interpretation of Signal (1543)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/13/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
A user facility reported, "the temperature is jumping from one temp to the next.Ex- 36 to 25 to 38 to 8." deroyal industries, inc.Requested return of the device but it has not yet been received.The investigation is not complete at this time.No further information is available at this time.We will provide follow-up report when additional information becomes available.
 
Event or Problem Description
A user facility reported, "the temperature is jumping from one temp to the next.Ex- 36 to 25 to 38 to 8.".
 
Additional Manufacturer Narrative
A user facility reported, "the temperature is jumping from one temp to the next.Ex- 36 to 25 to 38 to 8." deroyal industries, inc.Requested return of the device and it was received on 10/17/2025.As the lot was not reported, the specific device history record (dhr) was not able to be reviewed.Using samples returned from other similar complaints and the defective sample returned from this specific complaint, functional and dimensional testing was conducted that determined there was an instability at the connector side.Through further investigation, it was determined that a crimp set differed between the older molex machine and new automatic machine.Because of this difference, there was more pressure applied that caused the pin to deform in some cases.It was determined that this is what led to the connector instability.A complaint to sales ratio was conducted over the past two years and found to be (b)(4).Root cause: the root cause of the issue was the use of factory-installed crimp sets in the automatic machines with a pressing surface width narrower than the pin width.This dimensional incompatibility generated excessive localized mechanical stress during crimping, resulting in pin deformation, altered electrical contact geometry, and subsequent connector instability under temperature conditions.Immediate corrections: considering that the identified root cause was connector instability caused by insufficient electrical contact between the wire and the pin, all personnel involved were instructed to incorporate an additional verification step during functional testing.This step consists of gently manipulating the connector by moving it slightly once in an upward and downward direction and from side to side, in order to detect any potential intermittent or unstable conditions.Additionally, the sampling plan applied by quality control was escalated from normal inspection to tightened inspection.As a result, the sample size per box (2,000 units per box) was increased from 20 units to 32 units.This adjustment was implemented to strengthen detection capability, increase process oversight, and reduce the risk of escaping unstable units during the containment phase.Corrective actions: the automatic machine manufacturer (kingsing) and the local tool manufacturer (motrosa) were contacted.A dimensional modification of all automatic machine crimp sets was requested.Post implementation of these dimensional modifications indicate a substantial reduction in the instability rates.Deroyal will continue to monitor for trends around this item and issue.The investigation is complete at this time.No further information is available at this time.We will provide follow-up report if additional information becomes available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SKIN TEMPERATURE PROBE, 400 SERIES
Common Device Name
CONTINUOUS MEASUREMENT THERMOMETER
Manufacturer (Section D)
DEROYAL CIENTIFICA DE LATINOAMERICA S.R.L
global park, p.o. box 180-3006
602 parkway,
la aurora, heredia 146
CS  146
Manufacturer (Section G)
DEROYAL CIENTIFICA DE LATINOAMERICA S.R.L
global park, p.o. box 180-3006
602 parkway,
la aurora, heredia 146
CS   146
Manufacturer Contact
reagan facemire
200 debusk lane
powell, TN 37849
8653622106
MDR Report Key23416850
Report Number9613793-2025-00007
Device Sequence Number16060283
Product Code FLL
UDI-Device Identifier00749756046169
UDI-Public00749756046169
Combination Product (Y/N)N
Initial Reporter StateMN
Initial Reporter CountryUS
PMA/510(K) Number
K200631
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source User Facility
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 03/02/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number81-010400
Was Device Available for Evaluation? Device Returned to Manufacturer
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/16/2025
Supplement Date Received by ManufacturerNot provided
Initial Report FDA Received Date10/29/2025
Supplement Report FDA Received Date03/02/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-