AOMORI OLYMPUS CO., LTD. THUNDERBEAT OPEN FINE JAW TYPE X; ULTRASONIC SURGICAL DEVICE Back to Search Results
Model Number TB-0009OFX
Medical Device Problem Code Peeled/Delaminated (1454)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/02/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The device was returned to olympus for inspection.A root cause could not be identified.Based on the results of the investigation, it is likely the following led to the malfunction: the tissue pad in the grasping section was worn away, a part of the tissue pad was peeled away.There was a mark in the non-insulated area of grasping section which came in contact with the probe and was abraded against it.Should additional relevant information become available, a supplemental report will be submitted.Olympus will continue to monitor field performance for this device.
 
Event or Problem Description
It was observed that during the device evaluation, the thunderbeat exhibited peeling away of the tissue pad.There was no patient involvement.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
THUNDERBEAT OPEN FINE JAW TYPE X
Common Device Name
ULTRASONIC SURGICAL DEVICE
Manufacturer (Section D)
AOMORI OLYMPUS CO., LTD.
2-248-1 okkonoki
kuroishi-shi, aomori 036-0 357
JA  036-0357
Manufacturer (Section G)
AOMORI OLYMPUS CO., LTD.
2-248-1 okkonoki
kuroishi-shi, aomori 036-0 357
JA   036-0357
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key23417515
Report Number9614641-2025-01789
Device Sequence Number15699085
Product Code GEI
UDI-Device Identifier04953170394584
UDI-Public04953170394584
Combination Product (Y/N)N
Initial Reporter CountryKS
PMA/510(K) Number
K211838
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 10/29/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Model NumberTB-0009OFX
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/04/2025
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 10/02/2025
Initial Report FDA Received Date10/29/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured04/26/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
Yes
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-