| Model Number |
TB-0009OFX |
| Medical Device Problem Code |
Peeled/Delaminated (1454)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/02/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device was returned to olympus for inspection.A root cause could not be identified.Based on the results of the investigation, it is likely the following led to the malfunction: the tissue pad in the grasping section was worn away, a part of the tissue pad was peeled away.There was a mark in the non-insulated area of grasping section which came in contact with the probe and was abraded against it.Should additional relevant information become available, a supplemental report will be submitted.Olympus will continue to monitor field performance for this device.
|
| |
|
Event or Problem Description
|
|
It was observed that during the device evaluation, the thunderbeat exhibited peeling away of the tissue pad.There was no patient involvement.
|
| |
|