• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER BIOMET, INC. FEM SIZE E RIGHT; PROSTHESIS, KNEE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER BIOMET, INC. FEM SIZE E RIGHT; PROSTHESIS, KNEE Back to Search Results
Catalog Number 00588001502
Medical Device Problem Code Loosening of Implant Not Related to Bone-Ingrowth (4002)
Health Effect - Clinical Code Aseptic loosening (4883)
Date of Event 09/22/2025
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).D10: 00588001502 fem size e right lot# 65577833.00599003523 ng 15mm dist only fem augm sze lot# 64574554.00599003524 20mm distal only fem aug sz e lot# 63251030.00598802011 11mm dia 100mm length offset stem ext 145mm lot# 63988012.00545001831 trabecular metal femoral diaphyseal cone, 30mm, m, rt lot# 64800341.00588005012 12mm height size e articular surface with hinge post lot# 65532680.00801102024 allen medullary cement plugs 1- 24mm dia, flange/12 mm dia lot# 65626531.G2: australia.H6: suggested code (annex g)- mechanical (g04) - femur.The product will not be returned to zimmer biomet for investigation, as the device location is unknown.The investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.
 
Event or Problem Description
It was reported the patient was revised for femoral loosening two years, three months post implantation.Attempts have been made and no additional information has been provided.
 
Event or Problem Description
No further event information available at the time of this report.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional information.The following sections were updated: b1; b2; b5; b7; d2a; d2b; e1; g1; g3; h1; h2; h3; h4; h6.The following sections were updated: h6; h11.The reported event could not be confirmed due to lack of product.No product was returned.A photograph of the reported device was provided however the device failure could not be confirmed from the image.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.The device history record was reviewed and no discrepancies relevant to the reported event were found.Stem pn 00598802011, ln 63988012.Review of complaint history identified additional similar complaints for the reported item and no additional complaints for the reported part and lot combination.A definitive root cause cannot be determined.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FEM SIZE E RIGHT
Common Device Name
PROSTHESIS, KNEE
Manufacturer (Section D)
ZIMMER BIOMET, INC.
1800 w. center st.
warsaw IN 46580
Manufacturer (Section G)
ZIMMER BIOMET, INC.
1800 w. center st.
warsaw IN 46580
Manufacturer Contact
jennifer rapsavage
56 e. bell dr.
warsaw, IN 46582
5745260384
MDR Report Key23417828
Report Number0001822565-2025-03896
Device Sequence Number10300120
Product Code JWH
UDI-Device Identifier00889024199156
UDI-Public(01)00889024199156(17)271010(10)65577833
Combination Product (Y/N)N
Initial Reporter CountryAS
PMA/510(K) Number
K173057
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2023
Device Explanted Year2025
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Distributor
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 11/14/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Catalogue Number00588001502
Device Lot Number65577833
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/02/2025
Supplement Date Received by Manufacturer11/12/2025
Initial Report FDA Received Date10/29/2025
Supplement Report FDA Received Date11/14/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured10/11/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SEE H11 NARRATIVE.
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient SexMale
-
-