| Catalog Number |
81660095S |
| Medical Device Problem Code |
Use of Device Problem (1670)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/03/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Additional information has been requested and if received, will be provided in the supplemental report.
|
| |
|
Event or Problem Description
|
|
As reported: "when the nurse pulled to receive the unpacked gamma 4 u-lag screw, only the end cap detached and flew off, becoming contaminated.They opened a different size lag screw, substituted only the end cap, and completed the operation.".
|
| |
|
Event or Problem Description
|
|
As reported: "when the nurse pulled to receive the unpacked gamma 4 u-lag screw, only the end cap detached and flew off, becoming contaminated.They opened a different size lag screw, substituted only the end cap, and completed the operation.".
|
| |
|
Additional Manufacturer Narrative
|
|
The reported event could not be confirmed since the device was not returned for evaluation and no other evidences were provided.A review of the device history for the reported lot did not indicate any abnormalities.No corrective actions are required at this time.A review of the labeling did not indicate any abnormalities.No indications of material, manufacturing or design related problems were found during the investigation.More detailed information about the complaint event as well as the affected device must be available in order to determine the root cause of the complaint event.If the device is returned or if any additional information is provided, the investigation will be reassessed.
|
| |
|