SHIRAKAWA OLYMPUS CO., LTD. VISERA RHINO-LARYNGO VIDEOSCOPE Back to Search Results
Model Number ENF-V2
Medical Device Problem Codes Device Reprocessing Problem (1091); Poor Quality Image (1408)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/10/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The device was returned to olympus for inspection.A root cause could not be identified.Based on the results of the investigation, it is likely the following led to the malfunction: blurry screen when plugged in is due to due to dirty o.B lens, c0201 - electrical/electronic component problem identified, the performance of an electrical or electronic component was found to be inadequate.The most probable cause of this complaint is traced to d02 - cause traced to component failure, expected or random component failure without any design or manufacturing issue.Should additional relevant information become available, a supplemental report will be submitted.Olympus will continue to monitor field performance for this device.
 
Event or Problem Description
It was observed that during the device evaluation, the subject device exhibited a blurry screen when plugged in is due to due to dirty objective lens.There was no patient involvement.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VISERA RHINO-LARYNGO VIDEOSCOPE
Common Device Name
RHINO-LARYNGO VIDEOSCOPE
Manufacturer (Section D)
SHIRAKAWA OLYMPUS CO., LTD.
3-1 okamiyama
odakura, nishigo-mura,
nishishirakawa-gun, fukushima 961-8 061
JA  961-8061
Manufacturer (Section G)
SHIRAKAWA OLYMPUS CO., LTD.
3-1 okamiyama
odakura, nishigo-mura,
nishishirakawa-gun, fukushima 961-8 061
JA   961-8061
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key23418181
Report Number3002808148-2025-20171
Device Sequence Number8946052
Product Code EOB
UDI-Device Identifier04953170411342
UDI-Public04953170411342
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K061313
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 10/29/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Model NumberENF-V2
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/10/2025
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 10/10/2025
Initial Report FDA Received Date10/29/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured06/06/2013
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-