W. L. GORE & ASSOCIATES, INC. GORE® DRYSEAL FLEX INTRODUCER SHEATH; INTRODUCER, CATHETER
|
Back to Search Results |
|
| Catalog Number |
DSF1633 |
| Medical Device Problem Code |
Fluid/Blood Leak (1250)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/08/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
A1: no patient specific details have been provided.Therefore, the patient initials reflect the w.L.Gore internal case number.W.L.Gore & associates, inc.(gore) is submitting this report to comply with 21 cfr part 803, the medical device reporting regulation.This report is based upon information obtained by gore, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Blank fields present on this report include required fields and fields determined to be not applicable.Blank required fields indicate that the information was not provided, was deemed unavailable or was not applicable.This report does not constitute an admission or a conclusion by fda, gore, or its associates that the device, gore or its associates caused or contributed to the event described in the report.In particular, this report does not constitute a legal admission by anyone that the product described in this report has any defects or has malfunctioned, as defined from a legal standpoint.These words are included in the report and are fixed items for selection created by the fda, to categorize the type of event solely for the purpose of reporting pursuant to part 803.This statement should be included with any information or report disclosed to the public under the freedom of information act.
|
| |
|
Event or Problem Description
|
|
The following information was reported to gore: on (b)(6) 2025, the patient underwent an endovascular treatment for an abdominal aortic aneurysm using gore® dryseal flex introducer sheath as accessories.During use of the sheath, blood leakage was observed from the sheath valve.As the valve was not functioning properly, continued use was deemed difficult.The sheath was replaced with a new one, and the procedure was continued.No harm to the patient was reported.
|
| |
|
Additional Manufacturer Narrative
|
|
H6: a review of the manufacturing records for the device verified that the lot met all pre-release specifications.Product investigation report conclusion: the findings from the evaluation are consistent with the reported leakage, as water was observed leaking from the valve fill port stopcock.The cause for the stopcock leakage is attributed to a manufacturing deficiency of the supplied stopcock.A supplier corrective action report (scar) is currently open to investigate this failure, and this observation is within the scope of that scar.
|
| |
|