• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AVANOS MEDICAL INC. AVANOS MEDICAL; BALLARD CLOSED SUCTION CATHETER SYSTEM FOR NEONATES/PEDIATRICS, Y-ADAPTER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AVANOS MEDICAL INC. AVANOS MEDICAL; BALLARD CLOSED SUCTION CATHETER SYSTEM FOR NEONATES/PEDIATRICS, Y-ADAPTER Back to Search Results
Model Number 98383
Medical Device Problem Code Inadequacy of Device Shape and/or Size (1583)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/26/2025
Type of Reportable Event Malfunction
Event or Problem Description
Sunmed holdings llc.Received a single report that referenced two different incidences, which were associated with separate units, involving two different patients.This is the second of two reports.Refer to 8030647-2025-00112 for the first report.It was reported: the inline suction device could not pass through the opening of the connector, discovered during use on a ventilated infant.There was no reported injury.
 
Additional Manufacturer Narrative
The product involved in the report has not been returned for evaluation.A review of the device history record is in-progress.The udi-pi is not available as no product number was provided.All information reasonably known as of 29 oct 2025 has been included in this health authority report.Should additional information be obtained, a follow-up health authority report will be provided.The information provided by sun med holdings llc.Represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to sun med holdings llc.Sun med holdings llc.Has no independent knowledge of the event reported but is relaying the information that was provided by the user facility where the incident occurred.This product incident is documented in the sun med holdings complaint database and identified as (b)(4).This information is submitted pursuant to 21cfr803, in compliance with the medical device reporting requirement and should not be considered to be an admission that an sun med holdings llc.Product is defective or caused serious injury.
 
Event or Problem Description
Sunmed holdings llc.Received a single report that referenced two different incidences, which were associated with separate units, involving two different patients.This is the second of two reports.Refer to 8030647-2025-00112 for the first report.It was reported: the inline suction device could not pass through the opening of the connector, discovered during use on a ventilated infant.There was no reported injury.
 
Additional Manufacturer Narrative
Correction: d4; d5; g4 additional information-investigation completion: d4; h2; h6 the product involved in the report has not been returned for evaluation, additionally no photographic or videographic evidence was provided; therefore the complaint could not be confirmed.However, this is a known failure mode, the possible root cause was determined to be material segregated incorrectly.Operator retraining performed which included performing visual inspection every hour.The device history record for lot 1550819 was reviewed and the product was produced according to product specifications.All information reasonably known as of 19 dec 2025 has been included in this health authority report.Should additional information be obtained, a follow-up health authority report will be provided.The information provided by sun med holdings llc.Represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to sun med holdings llc.Sun med holdings llc.Has no independent knowledge of the event reported but is relaying the information that was provided by the user facility where the incident occurred.This product incident is documented in the sun med holdings complaint database and identified as (b)(4).This information is submitted pursuant to 21cfr803, in compliance with the medical device reporting requirement and should not be considered to be an admission that an sun med holdings llc.Product is defective or caused serious injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AVANOS MEDICAL
Common Device Name
BALLARD CLOSED SUCTION CATHETER SYSTEM FOR NEONATES/PEDIATRICS, Y-ADAPTER
Manufacturer (Section D)
AVANOS MEDICAL INC.
5405 windward parkway
alpharetta GA 30004
Manufacturer (Section G)
AVENT S DE RL DE CV
carretera internacional
salida norte no. 1053
magdalena, mx 84160
MX   84160
Manufacturer Contact
racheal cooper
2710 north ridge dr nw
ste a
grand rapids, MI 
MDR Report Key23420515
Report Number8030647-2025-00113
Device Sequence Number8178881
Product Code BSY
UDI-Device Identifier10609038983834
UDI-Public(01)10609038983834(10)30327230(17)290602(11)240612
Combination Product (Y/N)N
Initial Reporter CountryUK
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 12/19/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number98383
Device Catalogue Number195-5
Device Lot Number1550819
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/29/2025
Supplement Date Received by Manufacturer11/19/2025
Initial Report FDA Received Date10/29/2025
Supplement Report FDA Received Date12/19/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured11/12/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-