|
|
| Model Number |
SP-684 |
| Medical Device Problem Code |
Detachment of Device or Device Component (2907)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/10/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Event or Problem Description
|
|
It was reported that during a breast augmentation and reduction, during suturing or preparing the suture after patient incision, the needle detached from the suture.The needle did not fall into the patient cavity.An additional new suture was used without issue.There was no patient injury.
|
| |
|
Additional Manufacturer Narrative
|
|
Correction: e1 (phone number) additional information: g3, h3, h6 h3 evaluation summary: medtronic conducted an investigation based upon all information received.The device and a photo were available for evaluation.A visual inspection of the returned photo noted one needle, one blue suture and one retainer.The needle was noted to be disengaged from the suture.It was reported that during a breast augmentation and reduction, during suturing or preparing the suture after patient incision, the needle detached from the suture.The needle did not fall into the patient cavity.The reported issue was confirmed.The most likely cause was determined to be manufacturing related.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Search Alerts/Recalls
|
|
|