| Model Number |
4.75MM BC DL KNOTLESS CORKSCREW |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/21/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
On 10/21/2025, it was reported by a sales representative via (b)(4) that an ar-1943bc dl knotless corkscrew device anchor broke in half when implanting.This was discovered during the case with no patient harm.Additional information was received on 10/27/25 from rep that these were both discovered during separate rotator cuff repairs.The green punch that was provided was used to prepare the bone socket for insertion.The bone quality of the patients were good and all fragments were retrieved.The device was not able to be sent in.
|
| |
|
Additional Manufacturer Narrative
|
|
Investigation is in process.A follow-up report will be provided upon availability of additional information.
|
| |
|
Additional Manufacturer Narrative
|
|
Based on the information provided ¿ which may include the returned device (if available), photographs, videos, event description, and any additional field input ¿ arthrex has determined the most likely cause.The most likely cause is attributed to use-related factors, including: off-axis insertion, improper bone preparation, and prying or leveraging the device with excessive force.
|
| |
|