DEPUY ORTHOPAEDICS INC US TRI-LOCK BPS SZ 4 HI OFFSET; TRILOCK HIP STEM : HIP FEMORAL STEM Back to Search Results
Catalog Number 101214040
Medical Device Problem Code Use of Device Problem (1670)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 10/14/2025
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.H11 additional manufacturer narrative: d4: udi: as the lot number for the device involved in the event was not provided, the full udi is currently not available.
 
Event or Problem Description
It was reported that there was pinnacla inlay with massive wear due to malposition of the tri-lock stem.It possible to identify the head.
 
Additional Manufacturer Narrative
Product complaint # (b)(4) investigation summary pinnacla inlay with massive wear due to malposition of the tri-lock stem.Not possible to identify the head the product was not returned to depuy synthes; however, an x-ray and photos were provided for review.The photographs and x-ray attached were reviewed, however the reported "mispositioned" of the tri-lock bps sz 4 hi offset cannot be confirmed with the evidence provided.Therefore, the investigation could not draw any conclusions about the reported event.As the device was not returned, and as-received condition could not be assessed, a dimensional inspection and document/specification review were not completed.The overall complaint was not confirmed as the photographs and x-ray attached are insufficient to draw a conclusion about the reported event.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.There is no indication that a design or manufacturing issue has caused the complaint condition.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device history lot the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation (mre) was not performed.Added: d10 concomitant corrected: h4, h8.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRI-LOCK BPS SZ 4 HI OFFSET
Common Device Name
TRILOCK HIP STEM : HIP FEMORAL STEM
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
Manufacturer (Section G)
JTE WARSAW MFG SITE
700 orthopaedic drive
warsaw IN 46582
Manufacturer Contact
elba bello
700 orthopaedic dr.
warsaw, IN 46581
9083863534
MDR Report Key23423801
Report Number1818910-2025-18609
Device Sequence Number8242153
Product Code LPH
UDI-Device Identifier10603295001065
UDI-Public(01)10603295001065(17)220301(10)216562
Combination Product (Y/N)N
Initial Reporter CountrySZ
PMA/510(K) Number
P070026
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date03/01/2022
Device Catalogue Number101214040
Device Lot Number216562
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 10/13/2025
Supplement Date Received by Manufacturer02/03/2026
Initial Report FDA Received Date10/30/2025
Supplement Report FDA Received Date02/05/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured03/16/2012
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
PINN MAR NEUT 28IDX48OD
Outcome Attributed to Adverse Event Required Intervention;
Patient Age69 YR
Patient SexUnknown
-
-