| Model Number |
NA-U201H-8022 |
| Medical Device Problem Code |
Detachment of Device or Device Component (2907)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/07/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The investigation is ongoing.A supplemental report will be submitted when the investigation is completed or if additional information becomes available.
|
| |
|
Event or Problem Description
|
|
It was reported that the plastic tube of the aspiration needle had detatched from the handle.The reported issue occurred during an endoscopic ultrasound-guided fine needle aspiration (eus-fna) for a malignant lesion.The procedure was completed using another device and there were no reports of patient harm.
|
| |
|
Additional Manufacturer Narrative
|
|
The mdr supplemental report is being submitted to provide updated fields and final investigation results.Updated fields: d8, h2, h3, h6, h11.The device was returned for evaluation and the customer's allegation was not confirmed.A definitive root cause could not be identified.Based on the results of the investigation, it is likely the following led to the malfunction: no problem detected.Olympus will continue to monitor the performance of this device.
|
| |
|
Event or Problem Description
|
|
No additional information received from the customer.
|
| |
|