• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ETHICON INC. VCL CT BRD VIO 27IN 4-0 S/A RB-1; SUTURE, ABSORBABLE, SYNTHETIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ETHICON INC. VCL CT BRD VIO 27IN 4-0 S/A RB-1; SUTURE, ABSORBABLE, SYNTHETIC Back to Search Results
Catalog Number J304H
Medical Device Problem Codes Device Contaminated During Manufacture or Shipping (2969); Material Integrity Problem (2978)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Product complaint # (b)(4).Date sent to the fda: 10/29/2025.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.A manufacturing record evaluation was performed for the finished device lot, and no non-conformances were identified.Additional information was requested; the following was obtained: could you please elaborate further on the issue reported with the product? about two-thirds of the thread was white with some mottling.Do you have any photos available for visual analysis? no have please provide the source or name of person providing answers to follow-up questions:(b)(6).(b)(4) - event 1, (b)(4) - event 2.Could you please elaborate further on the issue reported with the product? unk do you have any photos available for visual analysis? yes, have.Could you kindly perform and document the follow-up attempt for the product return? furthermore, please ensure that the shipment tracking number is recorded in the notes or rmao sections.The device has been received at sukagawa and will be shipped.Please check rmao.Tracking number: (b)(4).Please provide the source or name of person providing answers to follow-up questions: (b)(6).To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.
 
Event or Problem Description
It was reported that a patient underwent an unknown procedure on an unknown date and suture was used.It was reported about two-thirds of the suture color was mixed with white in patches.Another product was used to complete the case.No sample will be returned.There were no adverse consequences to the patient.Additional information was requested.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VCL CT BRD VIO 27IN 4-0 S/A RB-1
Common Device Name
SUTURE, ABSORBABLE, SYNTHETIC
Manufacturer (Section D)
ETHICON INC.
1000 route 202
raritan NJ 08869
Manufacturer (Section G)
ETHICON INC.-JUAREZ
calle durango #2751
ciudad juarez 32575
MX   32575
Manufacturer Contact
orla o'mahony
1000 route 202
raritan, NJ 08869
3035526892
MDR Report Key23424677
Report Number2210968-2025-12306
Device Sequence Number15916235
Product Code GAM
UDI-Device Identifier10705031038615
UDI-Public(01)10705031038615(17)300131(10)106856
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K022269
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other
Type of Report Initial
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue NumberJ304H
Device Lot Number106856
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/08/2025
Initial Report FDA Received Date10/30/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured02/05/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-