• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ICU MEDICAL HEALTHCARE MANUFACTURING S.A. DE C.V LEVEL 1 FAST FLOW FLUID WARMER; WARMER, BLOOD, NON-ELECTROMAGNETIC RADIATION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ICU MEDICAL HEALTHCARE MANUFACTURING S.A. DE C.V LEVEL 1 FAST FLOW FLUID WARMER; WARMER, BLOOD, NON-ELECTROMAGNETIC RADIATION Back to Search Results
Catalog Number 8002950
Medical Device Problem Code Temperature Problem (3022)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/10/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional information becomes available.
 
Event or Problem Description
It was reported that the device overshot the targeted temperature and then cooled down very slowly.The device did not reach or maintain the desired temperature within 20 minutes of idle time after the procedure.There was no patient involvement, no injury was reported.
 
Additional Manufacturer Narrative
No product was returned.We are unable to confirm the reported complaint.If the product is returned, the manufacturer will reopen this complaint for further investigation.The root cause was unable to be established.The service history review had no indication that the complaint was related to a service of the device within the review period.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LEVEL 1 FAST FLOW FLUID WARMER
Common Device Name
WARMER, BLOOD, NON-ELECTROMAGNETIC RADIATION
Manufacturer (Section D)
ICU MEDICAL HEALTHCARE MANUFACTURING S.A. DE C.V
carretera miguel aleman
km 21.7
apodaca CP666 03
MX  CP66603
Manufacturer Contact
reed covert
6000 nathan lane n
minneapolis, MN 55442
2247062300
MDR Report Key23424869
Report Number3012307300-2025-11934
Device Sequence Number9863533
Product Code BSB
UDI-Device Identifier50695085829506
UDI-Public(01)50695085829506(17)151223(10)S103B02310
Combination Product (Y/N)N
Initial Reporter StateVA
Initial Reporter CountryUS
PMA/510(K) Number
BK020043
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source User Facility
Initial Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup
Report Date (Section B) 12/10/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number8002950
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/10/2025
Supplement Date Received by Manufacturer11/26/2025
Initial Report FDA Received Date10/30/2025
Supplement Report FDA Received Date12/10/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured12/23/2015
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-