(b)(4).Date sent: 10/30/2025.D4: batch #163c08.Investigation summary: the product was returned for evaluation.Visual inspection and functional testing were conducted on the returned device.Visual analysis of the returned sample determined that one ecr60w reload was received unfired with the one-piece sled missing, the reload pan was totally detached and bent.In addition, 2 loose single drivers and 1 loose double driver making the reload non-functional.No functional test was performed due to the condition of the reload.It is possible that when removing the staple retainer in an upward and distal to proximal sliding manner prior to loading the reload into the device can eject the sled from the proximal end of the reload.The ifu instruct the user to leave the staple retainer in position until the reload is loaded into the device.It is possible that handling by the customer could dislodge the pan.Dislodgement as a result of the removal of the reload from the device is the most likely scenario.As part of ethicon endo surgery¿s quality process, all devices are manufactured, inspected, and released to approved specifications.A manufacturing record evaluation was performed for the finished device ecr60w with lot number 225c23; and no non-conformances were identified.A manufacturing record evaluation was performed for the finished device batch number 163c08, and no non-conformances were identified.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|