| Catalog Number |
4825202 |
| Medical Device Problem Codes |
Adverse Event Without Identified Device or Use Problem (2993); Insufficient Device Problem Information (3190)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/04/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Investigation: the run data file (rdf) was analyzed for this event.The analysis of the run data file did not identify a conclusive root cause for the higher-than-expected wbc content reported in the rbc product for this collection.No unusual process variable was identified from the signals in the run data file.Based on the available information, it cannot be ruled out that the loading of the disposable set, particularly the rbc filter, may have contributed to the higher-than-expected wbc content measured in the rbc product.Based on the available information, it also cannot be ruled out that the higher-than-expected wbc content in the rbc product may have been donor-related.Investigation is in process, a follow-up report will be provided.
|
| |
|
Event or Problem Description
|
|
The customer would like the run data file investigated to determine a possible cause for the elevated white blood cell (wbc) content in the red blood cell product (rbc).Donor unit #: (b)(6) there was not a transfusion recipient or patient involved at the time of the residual wbc testing, therefore no patient information is reasonably known at the time of the event.The platelet collection set is not available for return because it was discarded by the customer.
|
| |
|
Additional Manufacturer Narrative
|
|
This report is being filed to provide additional information in d.4, h.6 and h.11.Investigation: the run data file (rdf) was analyzed for this event.The analysis of the run data file did not identify a conclusive root cause for the higher-than-expected wbc content reported in the rbc product for this collection.No unusual process variable was identified from the signals in the run data file.Based on the available information, it cannot be ruled out that the loading of the disposable set, particularly the rbc filter, may have contributed to the higher-than-expected wbc content measured in the rbc product.Based on the available information, it also cannot be ruled out that the higher-than-expected wbc content in the rbc product may have been donor-related.The device history records (dhr) were reviewed for this lot.There were no events noted in the dhr that would have contributed to the elevated wbc count experienced by the customer.A disposable complaint history search was performed for this lot and found 1 other report for a similar issue on this lot.See mdr: 1722028-2025-00242 root cause: a root cause assessment was performed for this complaint.The analysis of the run data file did not identify a conclusive root cause for the higher-than-expected wbc content reported in the rbc product for this collection.No unusual process variable was identified from the signals in the run data file.Based on the available information, it cannot be ruled out that the loading of the disposable set, particularly the rbc filter, may have contributed to the higher-than-expected wbc content measured in the rbc product.Based on the available information, it also cannot be ruled out that the higher-than-expected wbc content in the rbc product may have been donor-related.
|
| |
|
Event or Problem Description
|
|
The customer would like the run data file investigated to determine a possible cause for the elevated white blood cell (wbc) content in the red blood cell product (rbc).Donor unit #: (b)(6) there was not a transfusion recipient or patient involved at the time of the residual wbc testing; therefore no patient information is reasonably known at the time of the event.The platelet collection set is not available for return because it was discarded by the customer.
|
| |
|
Search Alerts/Recalls
|