|
MENTOR TEXAS MENTOR MEMORYGEL BREAST IMPLANT; PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED
|
Back to Search Results |
|
| Catalog Number |
3507400BC |
| Medical Device Problem Code |
Material Rupture (1546)
|
| Health Effect - Clinical Code |
Deformity/ Disfigurement (2360)
|
| Date of Event |
08/13/2025
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that a 68-year-old caucasian female patient underwent a primary breast augmentation with a 400cc mentor memorygel breast implant and experienced a rupture on the right side post-operatively, which was diagnosed with an office visit.As a result, the patient underwent explantation on (b)(6) 2025 and replaced with non-mentor devices (allergan).
|
| |
|
Additional Manufacturer Narrative
|
|
Since the device has not been returned for analysis, no product failure analysis can be conducted, and no determination of possible contributing factors can be made.As such, the investigation will be closed.If the complaint device is received in the future, the investigation will be reopened and conducted as appropriate.A manufacturing record evaluation is in progress.Once completed, a supplemental report will be submitted.Reason for device explant and/or reoperation: rupture.D4: udi: the product made prior to udi compliance date.Mentor is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which mentor has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, mentor, or its employees that the report constitutes an admission that the device, mentor, or its employees caused or contributed to the potential event described in this report.Manufacturer¿s reference number: (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
On november 24, 2025, the mentor failure analysis lab has received the device for evaluation.On november 27, 2025, the evaluation for the device was completed.Device evaluation summary: visual analysis of the returned device revealed that the breast implant was ruptured and received in two parts.In addition, shell abrasion was observed at the edge of the rupture.These findings are consistent with normal wear of the device.Shell abrasion suggests in-vivo folding or creasing of the device, which may be caused by continuous and sustained stresses to the device, such as an excessively small breast pocket or the folding or wrinkling of the shell within the breast pocket.In some cases, breast implants may also experience normal wear over time.As part of mentor's quality process, all devices are manufactured, inspected, and released according to approved specifications.Based on the results of this investigation, it was determined that the product met the manufacturing release criteria, and that normal wear may have caused the observed defect.No corrective and preventive action (capa) is required now.Additional complaint information monitoring for potential safety signals will be conducted through complaint trending as part of post market surveillance.A manufacturing record evaluation (mre) was performed, and no anomalies were found related to this complaint.In addition, the mre verifies that the device was manufactured in accordance with documented specification and procedures.Manufacturer¿s reference number: (b)(4).
|
| |
|
Search Alerts/Recalls
|
|
|