| Catalog Number |
0165SI12 |
| Medical Device Problem Code |
Deflation Problem (1149)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/17/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The investigation is still in progress.Once the investigation is complete, a supplemental report will be filed.Udi format updated with available product information per gudid.H11: sections a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
| |
|
Event or Problem Description
|
|
It was reported that during pretesting, it took a considerable amount of time to drain the water from the foley catheter.All was drained after 30 minutes was spent.Medicon syringe was used.
|
| |
|
Additional Manufacturer Narrative
|
|
The reported event could not be reproduced during investigation, and the returned device functioned as intended.While the event remained unconfirmed, the issue is understood to be associated with syringe engagement, and there is no evidence of catheter failure.Capa # 13490418 was opened to document the investigation.The labeling and instructions for use were found to be adequate.A review of the device history record did not show any problems or conditions that would have contributed to the reported issue.Therefore, no additional action is required at this time.Corrections: d, h.Udi format updated with available product information per gudid.H11: sections a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
| |
|
Event or Problem Description
|
|
It was reported that during pretesting, it took a considerable amount of time to drain the water from the foley catheter.All was drained after 30 minutes was spent.Medicon syringe was used.
|
| |
|
Search Alerts/Recalls
|