• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ARTHREX, INC. DISSECTOR, SJ 3.0MM X 7CM; SAW BLADE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ARTHREX, INC. DISSECTOR, SJ 3.0MM X 7CM; SAW BLADE Back to Search Results
Model Number DISSECTOR, SJ 3.0MM X 7CM
Medical Device Problem Code Material Discolored (1170)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/30/2025
Type of Reportable Event Malfunction
Event or Problem Description
On 09/30/2025, it was reported by a sales representative via (b)(4) that an ar-7300ds dissector has wear marks at the shp connection.The photo provided by the sales rep showed discoloration along the shaft of the dissector.Per the sales rep, the other devices being used with the dissector were ar-8330h shaver handpiece, an ar-8305 synergy shaver console, and an ar-8310 shaver footswitch.This was discovered during a case with no patient involvement.Additional information was received via medwatch on 10/10/2025: a facility representative reported that an ar-8305 synergy resection shaver console was used during a procedure on (b)(6) 2025.After the surgeon removed the arthroscopic shaver from the patient and was about to place it on the mayo stand, the surgical team observed that the device remained active despite the switch being turned off.Additional information was received via phone on 10/15/2025: per the rep, the event occurred outside the patient, on the mayo stand, after the ar-7300ds had been used and was no longer required for the procedure.It is unknown how long the ar-7300ds was used continuously, and whether any wear marks were present at the proximal end where it connects to the ar-8330h shaver handpiece.No adverse effects were reported during or after the surgery.Due to this issue, there were no injuries to the patient, surgeon, or staff.The incident occurred following a lateral ankle reconstruction procedure.The case was completed successfully without any delays or additional anesthesia.
 
Additional Manufacturer Narrative
Investigation is in process.A follow-up report will be provided upon availability of additional information.
 
Additional Manufacturer Narrative
Additional information: g3, h3, h6.Complaint allegation is confirmed by the attached picture, which shows outer shaft exhibiting damage caused by thermal exposure/overheating.Based on the information provided which may include the device (if available and returned), pictures, videos, event description, and any additional information from the field, arthrex was able to conclude a most likely cause.The most likely cause for the reported failure can be attributed to user error due to running the device without the flow of irrigation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DISSECTOR, SJ 3.0MM X 7CM
Common Device Name
SAW BLADE
Manufacturer (Section D)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer (Section G)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer Contact
jared engle
8009337001
MDR Report Key23429804
Report Number1220246-2025-04917
Device Sequence Number15523962
Product Code GFA
UDI-Device Identifier00888867041943
UDI-Public(01)00888867041943(11)250107(17)291231(10)15369249
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other
Initial Reporter Occupation Administrator/Supervisor
Type of Report Initial,Followup
Report Date (Section B) 01/30/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberDISSECTOR, SJ 3.0MM X 7CM
Device Catalogue NumberAR-7300DS
Device Lot Number15369249
Was Device Available for Evaluation? Yes
Distributor Facility Aware Date09/30/2025
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/30/2025
Supplement Date Received by Manufacturer01/29/2026
Initial Report FDA Received Date10/30/2025
Supplement Report FDA Received Date01/30/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured01/07/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-