MEDTRONIC PUERTO RICO VILLALBA SPRINT QUATTRO SECURE MRI SURESCAN; DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER Back to Search Results
Model Number 694758
Medical Device Problem Codes Fracture (1260); High impedance (1291); Impedance Problem (2950)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/03/2025
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Event or Problem Description
It was reported that alerts triggered for high undefined impedance on both of the right ventricular (rv) lead's defibrillation coils.Impedance on both coils as well as pacing impedance was noted to be erratic.One of the coils was reported to have cracked.One of the coils was turned off and the lead was capped and replaced.It was also reported that the left ventricular (lv) lead exhibited high erratic impedance when configured with one of the coils.The lv lead was replaced during the procedure as the physician felt the lead placement was suboptimal though the lead was also reported to have non-specifically deteriorated.No patient complications have been reported as a result of this event.
 
Additional Manufacturer Narrative
Product event summary: the lead was not returned for analysis, however, performance data collected from the device was received and analyzed.Analysis of the device memory indicated the impedance of the right ventricular defibrillation coil was beyond the expected upper range.Analysis of the device memory indicated the impedance trend of the right ventricular defibrillation coil was variable.Analysis of the device memory indicated the impedance of the superior vena cava defibrillation coil was beyond the expected upper range.Analysis of the device memory indicated the impedance trend of the superior vena cava defibrillation coil was variable.Analysis of the device memory indicated the impedance trend of the right ventricular pacing lead was variable.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Additional Manufacturer Narrative
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Event or Problem Description
It was further reported that the physician confirmed that the rv coil had high impedances as the svc coil had been turned off and the rv coil continued to have high impedances.The physician also confirmed that there was nothing mechanically wrong with the lead; it was simply placed in the coronary sinus body anterior lying in a possible diagonal vein and thought a different lead and position would work better physiologically for the patient.
 
Event or Problem Description
It was further reported by the patient that their previously replaced rv and lv leads had corroded.The patient was now experiencing increased heart failure symptoms.The physician believed that the newly placed left ventricular (lv) lead was not positioned for "true left bundle branch block (lbbb)".It was additionally noted that the previously capped rv lead experienced a confirmed fracture.A second revision procedure was performed to explant the previously capped rv lead, and explant and replace the recently implanted lv lead.No further patient complications have been reported as a result of this event.
 
Additional Manufacturer Narrative
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Additional Manufacturer Narrative
Product event summary: the distal portion of the lead was returned and analyzed.The right ventricular defibrillation coil conductor was fractured in-vivo in the anchoring sleeve area.The internal rv defib conductor was fractured at 23.5 cm.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SPRINT QUATTRO SECURE MRI SURESCAN
Common Device Name
DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER
Manufacturer (Section D)
MEDTRONIC PUERTO RICO VILLALBA
road 149 km 56.3
villalba PR 00766
Manufacturer (Section G)
MEDTRONIC PUERTO RICO VILLALBA
road 149 km 56.3
villalba PR 00766
Manufacturer Contact
gerard torenvliet
8200 coral sea st ne
mounds view, MN 55112-4391
7635269277
MDR Report Key23429971
Report Number2649622-2025-30544
Device Sequence Number9798262
Product Code LWS
UDI-Device Identifier00643169692374
UDI-Public(01)00643169692374(17)200914
Combination Product (Y/N)N
Initial Reporter StateMI
Initial Reporter CountryUS
PMA/510(K) Number
P920015
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2019
Device Explanted Year2026
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup,Followup,Followup
Report Date (Section B) 03/18/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date09/14/2020
Device Model Number694758
Device Catalogue Number694758
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/03/2025
Supplement Date Received by Manufacturer11/03/2025
12/15/2025
01/27/2026
03/03/2026
Initial Report FDA Received Date10/30/2025
Supplement Report FDA Received Date11/20/2025
01/02/2026
02/13/2026
03/18/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured10/04/2018
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
DTMA1Q1 CRT-D, 6725 ADAPTOR, 429888 LEAD
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Age80 YR
Patient SexFemale
Patient Weight73 KG
-
-